-34, "Not allowed command under cal"
Under calibration this command does not available.
7-4. Calibration Error
If you do calibration, Read back calibration inside the product will be proceeded.
-74, "ADC-V low limit over"
Voltage value is over from ADC-V low limit.
-75, "ADC-V high limit over"
Voltage value is over from ADC-V high limit.
-76, "ADC-A low limit over"
Current value is over from ADC Low limit.
-77, "ADC-A high limit over"
Current value is over from ADC High limit.
Error will be checked and saved in non-volatile memory.
If error occurs, error lamp will be on.
-80, "Memory limit volt error"
There's an error in voltage limit value.
-81, "Memory limit curr error"
There's an error in current limit value.
-82, "Memory max volt error"
There's an error in max voltage value.
-83, "Memory max curr error"
There's an error in max current value.
-84, "Memory volt decimal error"
There's an error in voltage value's decimal.
-85, "Memory curr decimal error"
There's an error in current value's decimal.
-86, "Memory volt length error"
There's an error in voltage value length (Digit).
- 91 -