
21 Appendix
70 / 80
Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
Warning
The use of accessories, transducers and cables not specified by OCULUS
(for example as replacement parts) may result in increased emissions or
decreased immunity of the Corvis® ST.
Use only the original accessories, transducers and cables specified by
OCULUS.
The use of accessories, transducers and cables specified by OCULUS with
devices other then the Corvis® ST may result in increased emissions or
decreased immunity of the other device.
Do not use the accessories, transducers and cables specified by
OCULUS with devices other then the Corvis® ST.
To be in compliance with the requirements of the IEC 60601-1-2 6.1 and
6.2 the following types of equipment, accessories, power adapters and ca-
bles must be used.
Order number
De
s
cription
72100
72200
Corvis® ST
05200905
Cable with plug, EU standard
1.8 m
05200910
Cable with plug, US standard
1.8 m
05200915
Cable with plug, GB standard
1.8 m
05200920
Cable with plug, AU standard
1.8 m
05200601
USB cable
015692000010
USB FS MED Isolator
05150285
Power Supply
15 V DC; 6 A