Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
Do not force any electrical connections. If you are unable to connect
a plug, check whether the plug fits the socket.
If you detect any damage to the connection, let our service personnel
repair the defect.
Establish an USB connection only with the OCULUS USB FS
MEDIsolator (Nr. 01 56920 00 010).
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Do not use the Corvis® ST with wireless technology, for example with
wireless USB.
Data responsibility:
The device itself is not designed to connect with
the internet, but only with a PC. It does not require the internet to
function.
Do not connect with the internet while using the device. It is consid-
ered misuse.
If you elect to connect the PC to the internet for other purposes you
are responsible for ensuring data security.
Patient environment information
Patient environment is the area where patients can come into contact
with any part of a medical electrical equipment (ME equipment) or with
another person being in contact with the ME equipment.