Instruction Manual Corvis® ST (G/72100/XXXX/EN 1019 Rev01)
Information about operation
Never put a damaged Corvis® ST into operation.
Only operate the Corvis® ST using original accessory parts supplied
by us, and only when the device is in technically correct working
order.
Before first use: Let OCULUS or an authorized dealer train you in the
operation of the Corvis® ST.
Only operate the device if you have understood the instruction
manuals.
Do not put the Corvis® ST, including the rechargeable battery or
cable, down onto devices that produce heat, heaters (e.g. radiators)
microwaves or similar.
Do not put any heavy objects onto the unit or the cables.
Information for Maintenance
Germany: The operating company must ensure that the device
undergoes technical measurement testing every 2 years according to
MPBtreibV, Appendix 2 Tonometer.
Only authorized persons are allowed to perform the software
updates.
To ensure that it functions correctly and safely we recommend the
following:
Have the Corvis® ST checked every two years by our service
department or an authorized dealer. If an error occurs which you are
unable to correct, label the Corvis® ST as "out of order" and contact
our service department or an authorized dealer.
Information for removal and disposal
When you disconnect electrical connections, pull on the respective
connectors, not the cord.
Dispose of the device according to legal regulations.
Instructions on Electrical Safety
Attention
Risk of personal injury or damage to property due to an incorrect level of
safety
Connecting the Corvis® ST with its non-medical electrical equipment (e.g.
data processing equipment) to a medical electrical system must not result
in a patient safety level below that prescribed by DIN EN ISO 60601-1. If
making this connection leads to the leakage current threshold being ex-
ceeded, protective measures including a circuit breaker must be in place.