NOTE: When using high pressure oxygen sources, all of the functional oxygen controls may be
utilized, e.g. auxiliary oxygen selector, oxygen flowmeter, oxygen flush valve, and oxygen
selector to power the ventilator.
NOTE: When using an oxygen concentrator as an oxygen source, there is ONLY enough
oxygen pressure to power the oxygen flowmeter. No other oxygen-powered/related systems
will be able to function. The O
2
flush will be powered by the backup "D" or "E" oxygen cylinder
when the low pressure O
2
concentrator is selected for use.
WARNING: If the O
2
flush button is depressed when using an O
2
concentrator, the flush
valve will exhaust the oxygen concentrator reservoir for at least 10 seconds or longer. Be
certain that the backup O
2
cylinder is turned on.
WARNING: If the oxygen concentrator is employed and the operator desires to utilize the
mechanical ventilator, compressed air is the only power source that can energize the ventila-
tor for operation. Once high-pressure oxygen is again available, the O
2
pressure toggle switch
can be returned to the high-pressure position and all O
2
-related systems will
then operate normally.
5. Flush Button: Press the button inwards to activate 100 % oxygen through the system.
When activated, the oxygen gas flows from the flush button to the Common Gas Outlet and
thence directly to the inspiratory side of the patient tubing circuit.
NOTE: One way valves prevent the oxygen flow from the Flush Button from mixing with other
gases and vapors already in use from the flowmeters and vaporizer.
6. Flowmeters: The oxygen and air flowmeters are operated by turning the control knobs
clock-wise for decreased flow and counter-clockwise for increased flow.
NOTE: Titration of FIO2 is managed by using both flowmeters and adjusting them until the
desired FIO2 is observed on the O
2
monitor read-out.
7. Ventilator Power Selector Switch: This toggle has three positions:
A. Center-OFF
B. Right-Oxygen
C. Left-Air
NOTE: Selection of gas power from the Air Compressor (if available) will allow O
2
sources to
be conserved.
WARNING: The comppac compressor module is the only recommended compressor for use
with the Magellan-2200 Model-2 air systems. Non-Medical Grade air compressor-output
may include oil which can cause physiologic problems for the Patient and could result in a
combustible O
2
-oil mixture or, damage the equipment beyond reasonable repair.
8. Common Gas Outlet: Provides a channel for all low flow gas and vapor to flow to the
inspiratory side of the CO
2
Absorber and then on to the patient and the rest of the circle system.
NOTE: An adapter is available to insert into the CGO in order that the operator may use
either a Jackson-Reese or Mapleson "D" system.
Vaporizer Mounting and General Use
1. Dismounting:
A.Remove the two holding knob-screws holding the retainer plate located on the
bottom-front of the vaporizer. Then remove the retainer plate.
B. Remove the two holding knobs and spacer plate from the back plate located on the
backside of the frame behind the vaporizer.
Содержание Magellan-2200
Страница 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Страница 26: ...26...
Страница 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Страница 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Страница 44: ...44 GAS FLOW DIAGRAM III Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch on BAG Position...
Страница 45: ...45 GAS FLOW DIAGRAM IV Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch o Vent Position...
Страница 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...
Страница 51: ...51 MAGELLAN 2200 ANESTHESIA MACHINE FRONT...
Страница 52: ...52 MAGELLAN 2200 ANESTHESIA MACHINE BACK...