6. CO2 Absorber Canister:
Attach the CO
2
Absorber Canister to the Absorber Support Arm.
BAG AND VENT ARM TUBING
7. Attach the
RED
labeled tubing to the outlet of the ventilator as shown on the label. Then
attach the other end of the tubing to the ventilator inlet port of the bellows (far right port)
using the 17mm connector.
8. Attach the
GREEN
labeled tubing to the brass outlet port of the B/V Switch and the
opposite end to the bottom right connector ("L" shaped) located on the bottom-right
position of the CO2Absorber.
9. Attach the
BLUE
labeled tubing to the right side inlet port of the
Bellows (labeled blue) and the opposite end to the lower outlet port of the
B/V Switch (labeled "To Bellows").
10. Patient Tubing Circuit:
Attach the preferred Patient Tubing Circuit to the CO
2
Absorber inlet and outlet
connecting points.
NOTE: Any patient tubing circuit approved for use with an anesthesia machine may be
used. The factory supplies a King Systems F-360-61 circuit with each machine as it is initially
shipped.
Содержание Magellan-2200
Страница 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Страница 26: ...26...
Страница 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Страница 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Страница 44: ...44 GAS FLOW DIAGRAM III Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch on BAG Position...
Страница 45: ...45 GAS FLOW DIAGRAM IV Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch o Vent Position...
Страница 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...
Страница 51: ...51 MAGELLAN 2200 ANESTHESIA MACHINE FRONT...
Страница 52: ...52 MAGELLAN 2200 ANESTHESIA MACHINE BACK...