(a) Summary Statement
The comppac OEM ventilator is a gas powered automatic ventilator intended to provide
ventilation during anesthesia, in a military field hospital, to adult and paediatric patients
greater than 20 kg (44 lb). The comppac OEM ventilator comprises two separate cases, one
case containing the ventilator that is fixed into an Oceanic Magellan 2200 portable
anaesthetic machine and the other case containing the compressor which is free standing
and provides air for both the ventilator and anaesthetic machine flowmeter.
The primary role of the comppac OEM ventilator is to provide patient ventilation when
fitted to the Oceanic Magellan 2200 Portable Anaesthesia Machine. It can be powered by
compressor oxygen, compressed air or by an electrically operated compressor. The
ventilator can be driven in various ways:
i) Oxygen from a nominal 60psi supply which can be powered by;
• Regulated compressed cylinder, or
• Wall-mounted outlets
ii) Air from a nominal 60psi supply which can be powered by:
• Regulated compressed cylinder, or
• Wall-mounted outlets
iii) Internal compressor providing compressed, fresh air. This can be powered by:
• internal 24V battery, or
• external 24/28V d.c. supply, or
• external 115V a.c. supply (via PS11 power supply), or
• external 240V a.c. supply (via PS11 power supply)
Graphical representation of electrical power source inputs, their relevant interfaces/
devices and outputs.
To run the comppac:
SECTION 1
The comppac OEM ventilator comprises of the following standard kit parts:
• Ventilator Module: Part No - 510A3053 (1)
• Compressor Module: Part No – 510A3060 (2)
• Rubber Boot: Part No – 510A2271 (3)
• NBC Filter: Part No – W7265 (4)
• Hose MA4 DISS to DISS: Part No – 500A4888 (5)
• 24/28 volt open-ended Vehicle Supply lead to comppac: Part No – 510A3096 (6)
• comppac ventilator / compressor User Manual: Part No – 504-2107 (Not shown)
• Water trap Assembly: Part No – 510A3105 (7)
• Interconnection cable: Part No – W288-001 (8)
• Reservoir Cover Spares: Part No – 510A2956 (Not shown)
• Sensing line: Part No – 510A3107 (9)
• NiCad Battery: Part No – 510-A1490 (10)
• Microbiological Filter: Part No – W8860 (11)
Figure 1: CompPac OEM System
— Charging While Operating V
entilator
Содержание Magellan-2200
Страница 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Страница 26: ...26...
Страница 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Страница 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Страница 44: ...44 GAS FLOW DIAGRAM III Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch on BAG Position...
Страница 45: ...45 GAS FLOW DIAGRAM IV Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch o Vent Position...
Страница 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...
Страница 51: ...51 MAGELLAN 2200 ANESTHESIA MACHINE FRONT...
Страница 52: ...52 MAGELLAN 2200 ANESTHESIA MACHINE BACK...