32
11. Perform Leak Check of the Breathing System
A. Set all gas flows to zero (or minimum).
B. Close PRV valve and occlude "Y" piece or install a reservoir bag to distal end of circuit.
C. Pressurize breathing system to about 30 cmH
2
O with O
2
Flush.
D. Ensure that pressure remains fixed for at least 10 seconds.
E. Open PRV valve and ensure that pressure decreases.
Manual and Automatic Ventilation Systems
12. Test Ventilation Systems and Unidirectional Valves
A. Place a second breathing bag on "Y" piece.
B. Set appropriate ventilator parameters for next patient.
C. Switch to automatic ventilation (ventilator) mode.
D. Fill bellows and breathing bag with O
2
Flush. Turn ventilator power switch to air or O
2
.
E. Set O2 flow to minimum, Air to zero.
F.
Verify that during inspiration, bellows deliver appropriate tidal volume and that during
expiration, bellows fills completely.
G. Set fresh gas flow to approx. 5 L/min.
H. Verify that the ventilator bellows and simulated lungs fill and empty approximately
without sustained pressure at end-expiration.
I.
Check for proper action of unidirectional valves located in CO
2
canister.
J.
Exercise breathing circuit accessories to ensure proper function.
K. Turn ventilator gas power switch to the "OFF" position and turn bag switch to
"BAG" mode.
L. Ventilate bag manually and assure inflation and deflation of artificial lung and
appropriate feel of system resistance and compliance.
M. Remove second bag from "Y" piece.
N. Turn B/V switch to "Vent". Allow bellows to fill.
O. Set minute volume contro to 8L/min and frequency to 10 B/min.
P.
Utilize a 1 liter breathing bag as a test lung.
Q. Turn on the ventilator to either air or O
2
depending on which gas you will be using to
drive the vent.
Содержание Magellan-2200
Страница 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...
Страница 26: ...26...
Страница 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...
Страница 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...
Страница 44: ...44 GAS FLOW DIAGRAM III Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch on BAG Position...
Страница 45: ...45 GAS FLOW DIAGRAM IV Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch o Vent Position...
Страница 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...
Страница 51: ...51 MAGELLAN 2200 ANESTHESIA MACHINE FRONT...
Страница 52: ...52 MAGELLAN 2200 ANESTHESIA MACHINE BACK...