Oceanic Magellan-2200 Скачать руководство пользователя страница 20

20

(e) Accessories
(i) Batteries
The comppac ventilator is designed to operate from the NiCad rechargeable version of a
widely stocked military 24V battery which was originally introduced to power the
“Clansman” communications pack; NATO No. 6140 - 99 - 620 - 8057. It is packaged in either
a plastic or metal casing but the comppac ventilator only accepts the plastic casing - there
are wider fixing lugs on the metal casing.

The Ni-Cad battery has an 85ºC internal thermal cut out.

WARNING: To avoid malfunction of the ventilator when using the battery as a power
source, periodically check the battery to ensure that there is sufficient charge to
power the ventilator. The battery may be trickle or fast charged using the PS11
power supply/charger unit (see section 2e) iii), however, this should be undertaken
in advance of use and, where necessary (ie: battery does not retain its charge), the
battery may need replacing.

This battery needs to be recharged within a month before use. Battery life will depend on
ventilator settings, the table below shows typical battery duration for given settings:

The batteries are field exchangeable.

CAUTION: It is recommended that the comppac ventilator is only used with a battery
installed, even if the ventilator is being powered by gas only, as the battery is used to
power the electronic alarm system. The ventilator will work without a battery or external
electrical supply, since it operates as a conventional gas powered device with the patient
protection devices and the high inflation pressure pneumatic alarm operating normally.
However, the user must be aware that in these circumstances, the electrically operated
alarms will NOT function.

(ii) Power Supply/Charger

WARNING: To avoid the potential risk of explosion associated with re charging the
Lithium version of battery, ensure that only a Ni Cad battery (Part No: 510-A1490/CE,
NATO No: 6140-99-620-8057) is installed. UNDER NO CIRCUMSTANCES SHOULD THE
LITHIUM BATTERY (NATO No: 6135-99-840-0109), OR ANY OTHER PRIMARY
BATTERY BE FITTED.

The Pneupac PS11 Mains Power Supply/Charging Unit is specially designed for use with
the comppac ventilator. Its 6 pin input socket will accept leads for connection to supplies
of 200/250V 50Hz, 100/120V 60Hz or 24/28V d.c. respectively. Each lead is wired to
connect to a different pair of pins.

The PS11/ comppac supply cable assembly (510A2421) connects between the 4-pin
socket on the power supply output and the Auxiliary Electrical Supply connector on the
panel of the comppac ventilator.

WARNING: To avoid the risk of fire or explosion or impaired ventilator performance,
always observe the correct polarity of the auxiliary supply. Ensure the supply is a
stable voltage of between 23 to 30 Volts capable of supplying 3 Amps (5 Amp surge
on motor start) and does not contain excessive interference.

When switched to ‘On & Trickle Chg’ the PS11 power supply provides the correct voltage
and current to drive the ventilator through one of the circuits and to simultaneously
trickle charge the internal battery through a second circuit. This trickle charging will
provide about 8 minutes running time for every hour of charging.

When the power supply is switched to the ‘Fast Charge’ position it will only energise the
third, fast charging, circuit. In this mode the rate of charge is increased by about 4 times
such that the internal battery will become fully charged in 4-5 hours. Fast charging will
only take place when the ‘On/Off’ switch on the panel of the ventilator is in the ‘Off’
position.

NOTE: Charging light may trip out due to type of battery and therefore may not
necessarily mean the battery is fully charged. It is best to trickle charge for longer peiords.

Alternatively, if the battery charging lead (510A2422) is connected to the output socket of
the power supply it can be used to fast charge a spare battery directly – outside the
ventilator.

Plug-in supply leads can be supplied to suit most mains outlet sockets used World-wide.

CAUTION: To avoid accidental disconnection of the power supply unit (PS11) when it is
connected to the comppac OEM unit, the mains lead should always be locked in position.

CAUTION: Avoid fast charging batteries outside the temperature range of +10ºC to +30ºC.
The batteries will not accept a full charge outside this range and, in certain circumstances,
the capacity of the battery may be permanently impaired. At low temperature, the
electrolyte may be electrolysed, and small quantities of hydrogen liberated, which in
extreme circumstances present an explosive risk.

An auxiliary power supply lead, part number 510A2582, is available from Smiths Medical
International Ltd. to which the user can fit specific connectors to suit alternative 24/28V
d.c. power supplies. This lead has only a single circuit so simultaneous battery charging
will not occur with its use.

(iii) NBC Filter

WARNING: When in use in a contaminated atmosphere, to avoid the risk of harm to
the patient, the NBC filter must be fitted.

The comppac ventilator is designed to accept any NBC filter with a NATO Stanag 4155

Содержание Magellan-2200

Страница 1: ...e Operators Manual 8005 Shannon Industrial Park Lane Atchison Kansas 66002 Tel 913 874 2000 Fax 913 874 2005 oceanicmedical lvnworth com CAUTION Federal law restricts this device to be used by or on t...

Страница 2: ...cessories 31 Post Use Checklist 36 Cleaning Disinfecting and Sterilization 37 Minor Maintenance Recommendations 37 Tools Needed for Maintenance 37 PROBLEM SOLVING GUIDE 38 VACUUM MANIFOLD 38 CYLINDER...

Страница 3: ...uld be under the constant surveillance of competent medical practitioners There is always the possibility of machine and alarm failure and some malfunctions require immediate corrective action 2 Vapor...

Страница 4: ...wer O2 flowmeter only Air Compressor May be used to power ventilator and air flowmeter Oxygen Analyzer Monitor OM 25 ME or equivalent Galvanic cell sensor life expectancy 2 years under normal conditio...

Страница 5: ...1 Removal and Mounting of Machine Turn on side open remove lid 2 Remove machine Turn open box so that wheels are down 3 Review parts in lid of ventilator box 5 Mount anesthesia machine to top of box s...

Страница 6: ...ing from the O2 concentrator should be connected to the specially labeled O2 concentrator only inlet nipple located above and to the right side of the Gas Manifold ASSEMBLY OF COMPONENTS FOR USE OF TH...

Страница 7: ...ttach the reservoir bag to bottom port of the scavenger Attach suction source tubing to the inlet port of scavenger control knob port NOTE In military field conditions the reservoir bag to the bottom...

Страница 8: ...ob screw from the back side of the frame to secure the Arm Ensure that the stem of the Arm is pointed upwards 2 Bag Vent Switch Assembly Arm Attach the Bag Vent Switch Assembly Arm to the top of the m...

Страница 9: ...rway Pressure Gauge Tubing connector to the bottom of the B V Switch Arm 5 Gas FlowTubing Attach the semi clear tubing from the Common Gas Outlet to the inlet connector of the CO2 absorber NOTE If the...

Страница 10: ...and the opposite end to the bottom right connector L shaped located on the bottom right position of the CO2Absorber 9 Attach the BLUE labeled tubing to the right side inlet port of the Bellows labeled...

Страница 11: ...et 2 Attach the blue O2 monitor sensor bracket into the Inspiratory outlet located on the left hand side of the front of the CO2 absorber 3 Attach the O2 monitor O2 sensor into the blue O2 sensor brac...

Страница 12: ...P USE and FUNCTIONAL CHECK 22 a Set Up 22 i comppac Ventilator 22 b Functional Check 24 SECTION 4 OPERATION 27 a User s Skill 27 b Setting of the Ventilator 27 i General 27 ii Ventilating the Patient...

Страница 13: ...nominal 60psi supply which can be powered by Regulated compressed cylinder or Wall mounted outlets iii Internal compressor providing compressed fresh air This can be powered by internal 24V battery o...

Страница 14: ...that could lead to death or serious injury to the user or patient that can occur if you do not obey all of the instructions given in this manual The following Warnings highlight aspects of the use of...

Страница 15: ...ained by Smiths Medical International Ltd to carry out such work Section 6c 21 WARNING Extreme environments may impair ventilator performance operator vigilance is required to monitor the patient 22 W...

Страница 16: ...associated with re charging the Lithium version of battery ensure that only a Ni Cad battery Part No 510 A1490 CE NATO No 6140 99 620 8057 is installed UNDER NO CIRCUMSTANCES SHOULD THE LITHIUM BATTER...

Страница 17: ...ropped to a pressure at which the ventilator will no longer be operating to specification if operating from a compressed gas supply With low pressure it shows red with adequate pressure it shows white...

Страница 18: ...the Magellan 2200 If the compressor module is used to supply the compressed air the supplied 1 5m hose and water trap must be fitted between the compressor module and Magellan 2200 air supply manifold...

Страница 19: ...o run the compressor The yellow LED is illuminated when the system has automatically chosen the battery to run the compressor Note Power for the alarm system is taken from the battery by default unles...

Страница 20: ...utput and the Auxiliary Electrical Supply connector on the panel of the comppac ventilator WARNING To avoid the risk of fire or explosion or impaired ventilator performance always observe the correct...

Страница 21: ...C Filter Canisters which are registered as NATO No 4240 99 132 0941 CAUTION The protective bung fitted to the NBC Filter should always be removed before fitting it to the comppac ventilator For use in...

Страница 22: ...specially threaded adapter in the recess of the rear face Screw it in fully until sealing contact is made with the sealing washer at the internal base of the adapter The anaesthetic machine bellows u...

Страница 23: ...23 MAGELLAN 2200 MODEL 2 MACHINE AIR COMPRESSOR AND POWER SUPPLY BACK VIEW...

Страница 24: ...of the ventilator and the ventilator alarms and compressor should start in the following sequence a The ventilator alarm LED s should illuminate in a clockwise sequence starting from the low power LED...

Страница 25: ...or The ventilator should always be run with a NBC filter in place as this reduces the operating noise as well as filtering the air entering the compressor Other than the protection provided to the pat...

Страница 26: ...26...

Страница 27: ...stantly monitor the patient pressure manometer during manual ventilation WARNING Always ensure that an alternative means of ventilation is available in the event of ventilator failure or malfunction a...

Страница 28: ...ons given in the functional check WARNING Failure to constantly monitor the patient whilst using this equipment may lead to death or serious injury The Inflation Pressure Monitor should be kept under...

Страница 29: ...ly trained personnel of the compressor reservoir are detailed in section 6 d iv Reassemble the system and carry out the function check specified in Section 3 v Stow the system in a clean area away fro...

Страница 30: ...at least 5 seconds E Open vaporizer and repeat C and D above F Remove suction bulb and reconnect fresh gas hose PRE USE CHECKLIST Emergency Ventilation Equipment 1 Verify backup ventilation equipment...

Страница 31: ...Adjust suction on and ensure bag collapses E Turn suction off Breathing System 9 Calibrate Oxygen Monitor A Ensure monitor reads 21 in room air B Verify low oxygen alarm is enabled and functioning C R...

Страница 32: ...ng inspiration bellows deliver appropriate tidal volume and that during expiration bellows fills completely G Set fresh gas flow to approx 5 L min H Verify that the ventilator bellows and simulated lu...

Страница 33: ...ful Center L Frame Controls and Gauges 1 Auxiliary O2 Flow Selector Designed to allow the operator to supply oxygen to a patient pre and post op with small bore tubing for a nasal cannula or disposabl...

Страница 34: ...he oxygen flow from the Flush Button from mixing with other gases and vapors already in use from the flowmeters and vaporizer 6 Flowmeters The oxygen and air flowmeters are operated by turning the con...

Страница 35: ...the filler 9 Loosen clamp screw and remove the key adapter 10 Insert the clamp screw and tighten clamp screw C Use During Procedure A Turn percent control knob to desired position B Adjust percent con...

Страница 36: ...PSI Optimal CAUTION If Oxygen is the desired gas pressure source ensure that the Gas PowerToggle Switch top left on L frame is in the High Pressure position NOTE Use of the Air Compressor is the prefe...

Страница 37: ...er Switch to either Air or O2 I Allow the ventilator to cycle at least twice J During the Inspiratory Phase of each cycle observe the depression level of the bellows K Adjust the MinuteVolume and or b...

Страница 38: ...Inspiratory Pressure is being generated O Check theTidalVolume with a spirometer and adjust the ventilator controls if necessary P Observe other monitoring devices if available for accuracy of respir...

Страница 39: ...turers directions for use Steam Sterilization The following components may be autoclaved at a temperature of 121 degrees C for fifteen minutes 1 Tubing and connectors from the common gas outlet to the...

Страница 40: ...not inflate Check Bag Vent Switch for vent position Check PRV Control Knob for closed position Check entire tubing circuit for leaks Oxygen Sensor does not calibrate Check O2 supply and O2 Pressure G...

Страница 41: ...Cylinders C E G H Size 1 Pressure 2200 psi 12 7 22 187 244 2 Factors duration 0 6 0 28 2 41 3 14 of flow Air 02 3 Formula duration of flow Flow minutes Cylinder press x factor liter flow Example E Cy...

Страница 42: ...42 GAS FLOW DIAGRAM I O2Flow from Gas Manifold to L Frame...

Страница 43: ...43 GAS FLOW DIAGRAM II Air Flow from Gas Mnifold to L Frame...

Страница 44: ...44 GAS FLOW DIAGRAM III Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch on BAG Position...

Страница 45: ...45 GAS FLOW DIAGRAM IV Air O2 From Flow Meters to Common Gas Outlet CGO Bag Vent Switch o Vent Position...

Страница 46: ...46 GAS FLOW DIAGRAM V Gas Flow From O2 Concentrator to O2 Flow Meter...

Страница 47: ...apter One each PEEP Valve and AccessoryTubing with Connector One each Military Accessory Bundle One each Respirometer with Adapters ADDITIONAL ITEMS NEEDED FOR OPERATION AND MAINTENANCE Note Military...

Страница 48: ...lowValve Bracket 12 18 8 32 Brass Nuts 12 19 Pressure Gauge Dual Scale Non Magnetic 12 20 Pressure Gauge Bracket 12 18 8 32 Brass Nuts 12 21 5 40 x Brass Nuts 12 22 Aluminum ExhalationValve Bracket 12...

Страница 49: ...mm Inlet Red ARMY BILL OF MATERIALS CONT D PART NO DESCRIPTION 21 0 Label L Frame 21 1 Label Vaporizer Warning 21 3 Label Ventilator Front Panel 21 4 Label Ventilator Side Panel 21 5 Label Magellan 22...

Страница 50: ...lters 3 and 3 metal spacers Tweezers for removing filters and spacers 1 16 Allen s wrench Squeeze bulb with CGO adapter Tubing color coded 3 and red adapter Instuctions for Service Sheet Krytox lubric...

Страница 51: ...51 MAGELLAN 2200 ANESTHESIA MACHINE FRONT...

Страница 52: ...52 MAGELLAN 2200 ANESTHESIA MACHINE BACK...

Отзывы: