8: Firmware and Advanced Communications
80
225-00000-000-11-201604
Byte Format
Byte 0
0x6C
Return Format (EP1In)
Byte 0
Byte 1
Byte 2
Read Result
ADC Value LSB
ADC Value MSB
If the operation was successful, the Read Result byte value will be 0x08. All other values
indicate the operation was unsuccessful.
Read Irradiance Factors
Reads 60 bytes of data, which is utilized for Irradiance Calibration information from the desired
EEPROM memory address.
Byte Format
Byte 0
Byte 1
Byte 2
0x6D
EEPROM Address LSB
EEPROM Address MSB
Return Byte Format
Byte 0
Byte 1
…
Byte 59
Byte 0
Byte 1
…
Byte 59
Write Irradiance Factors
Write 60 bytes of data, which is utilized for Irradiance Calibration information to the desired
EEPROM memory address.
Byte Format
Byte 0
Byte 1
Byte 2
Byte 3
…
Byte 62
0x6E
EEPROM Address LSB
EEPROM Address MSB
Byte 0
…
Byte 59
Query Status
Returns a packet of information containing the current operating information. Packet structure
is given below: