8: Firmware and Advanced Communications
225-00000-000-11-201604
103
Register
Address
Description
Default
Value
Min
Value
Max Value
Time Base
0x78
Offset Control
Bit 0 = Enable Auto-
Nulling
Bit 1 = Enable Auto-
Nulling Saturation
0
0
0xFFFF
N/A
0x7C
FPGA Programmed
(Read Only)
0x5501
N/A
N/A
N/A
0x80
Maximum Saturation
Level
0x55F0
0
0xFFFF
N/A
Notes: * - User should not change these values because spectrometer performance can be affected. This
information is included just for completeness
& - These values are controlled by other command interfaces to the Flame (i.e., Set integration
time command).
Read Register Information
Reads the values from any of the registers above. This command is sent to End Point 1 Out and
the data is retrieved through End Point 1 In.
Byte Format
Byte 0
Byte 1
0x6B
Register
Value
Return Format (EP1In)
Byte 0
Byte 1
Byte 2
Register Value
Value MSB
Value LSB
Read PCB Temperature
Reads the Printed Circuit Board temperature. The Flame contains a DS1721 temperature sensor
chip which is mounted to the underside of the PCB. This command is sent to End Point 1 Out
and the data is retrieved through End Point 1 In. The value returned is a signed 16-bit A/D
conversion value, which is equated to temperature by:
Temperature (
o
C) = .003906 * ADC Value
Содержание Flame-S
Страница 18: ...1 Introduction 8 225 00000 000 11 201604...
Страница 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Страница 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Страница 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Страница 58: ...6 Technical Specifications 48 225 00000 000 11 201604 Mechanical Diagram Figure 1 Flame Outer Dimensions...
Страница 74: ...7 Calibration 64 225 00000 000 11 201604...
Страница 130: ...Index 120 225 00000 000 11 201604...