5: How the Flame Spectrometer Works
225-00000-000-11-201604
39
3.
LongPass Absorbing Filter (optional):
If selected, an absorbing filter
is installed between the slit and the aperture in the SMA 905 bulkhead.
The filter is used to limit bandwidth of light entering spectrometer or to
balance color. Filters are installed permanently. A filter is for a specific slit.
If you anticipate needing the filter with multiple slit sizes, then you must
specify this at the time you order. You will know which filter is installed in each slit
because of the color-coded dots on the outside as shown in the figure and described in
the table below.
Item Code
Description
Dot 1
Dot 2
OF1-BG28
Bandpass filter, transmits >325 and <500 nm
blue
red
OF1-WG305
Longpass filter; transmits light >305 nm
black
white
OF1-U325C
Bandpass filter, transmits >245 and <390 nm
white
green
OF1-GG375
Longpass filter; transmits light >375 nm
red
black
OF1-GG395
Longpass filter; transmits light >395 nm
white
red
OF1-CGA420
Longpass filter; transmits light >420 nm
orange
white
OF1-GG475
Longpass filter; transmits light >475 nm
green
green
OF1-OG515
Longpass filter; transmits light >515 nm
pink
yellow
OF1-OG550
Longpass filter; transmits light >550 nm
orange
orange
OF1-OG590
Longpass filter; transmits light >590 nm
red
pink
OF1-RG695
Longpass filter; transmits light >695 nm
white
blue
OF1-RG830
Longpass filter; transmits light >830 nm
black
blue
OF1-CGA1000
Nonfluorescing longpass filter, transmits >1000 nm
red
green
OF1-CGA760
Nonfluorescing longpass filter, transmits >760 nm
blue
black
OF1-CGA780
Nonfluorescing longpass filter, transmits >780 nm
white
yellow
OF1-CGA830
Nonfluorescing longpass filter, transmits >830 nm
green
orange
OF1-CGA475
Nonfluorescing longpass filter, transmits >475 nm
yellow
pink
4.
Collimating Mirror (specify Standard or SAG+):
The collimating mirror is matched
to the 0.22 numerical aperture of our standard optical fibers. Light reflects from this
mirror, as a collimated beam, toward the grating. You can opt to install a standard
mirror or a NIR-enhancing but UV absorbing SAG+ mirror.
SAG+ mirrors are often specified for
fluorescence. These mirrors absorb nearly all UV
light, which reduces the effects of excitation scattering in fluorescence measurements.
Unlike typical silver-coated mirrors, the SAG+ mirrors won’t oxidize. They have excellent
reflectivity — more than 95% across the VIS-NIR.
Specify standard or SAG+ mirrors when ordering your spectrometer.
Содержание Flame-S
Страница 18: ...1 Introduction 8 225 00000 000 11 201604...
Страница 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Страница 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Страница 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Страница 58: ...6 Technical Specifications 48 225 00000 000 11 201604 Mechanical Diagram Figure 1 Flame Outer Dimensions...
Страница 74: ...7 Calibration 64 225 00000 000 11 201604...
Страница 130: ...Index 120 225 00000 000 11 201604...