8: Firmware and Advanced Communications
225-00000-000-11-201604
97
Packet #
End Point
# Bytes
Pixels
0
EP2In
64
0-31
1
EP2In
64
32-63
2
EP2In
64
64-95
…
EP2In
64
119
EP2In
64
3808
–3839
120
EP2In
1
Sync Packet
Packet 0
Byte 0
Byte 1
Byte 2
Byte 3
Pixel 0 LSB
Pixel 0 MSB
Pixel 1 LSB
Pixel 2 MSB
…
Byte 62
Byte 63
Pixel 31 LSB
Pixel 31 MSB
Packet 120
– Synchronization Packet (1 byte)
Byte 0
0x69
Set Trigger Mode
Sets the Flame Trigger mode to one of the following states. If an unacceptable value is passed,
then the trigger state is unchanged.
Data Value = 0
Normal (Free running) Mode
Data Value = 1
Software Trigger Mode
Data Value = 2
External Synchronization Trigger Mode
Data Value = 3
External Hardware Trigger Mode
Byte Format
Byte 0
Byte 1
Byte 2
0x0A
Data Value LSB
Data Value MSB
Query Number of Plug-in Accessories
Queries the number of Plug-in accessories preset. This is determined at power up and
whenever the Plug-in Detect command is issued.
Содержание Flame-S
Страница 18: ...1 Introduction 8 225 00000 000 11 201604...
Страница 32: ...2 Installation and Setup 22 225 00000 000 11 201604...
Страница 46: ...4 Troubleshooting 36 225 00000 000 11 201604...
Страница 54: ...5 How the Flame Spectrometer Works 44 225 00000 000 11 201604...
Страница 58: ...6 Technical Specifications 48 225 00000 000 11 201604 Mechanical Diagram Figure 1 Flame Outer Dimensions...
Страница 74: ...7 Calibration 64 225 00000 000 11 201604...
Страница 130: ...Index 120 225 00000 000 11 201604...