cautIon
movIng the table
dIrectIons for use
before doing any work on the
table, disconnect the power
cord from the power outlet
and secure the electrical
cord, and foot controls for
moving.
The table weighs as much
as 200 lbs.
Do Not
lift or
move the unit without using
proper safety procedures
or equipment.
Do Not
turn
the table on its side for any
reason. To do so could
cause damage to the base
and
will
void the warranty.
There are no serviceable
items that require this action
.
Never lift the table by grasping
the table top as the table
top is not
designed to hold
the weight of the base and
you could damage the table.
at low height range
The table weighs approximately 200 lbs. You will need 4 people working
together if it is necessary to move the table. Never lift the table by grasping
the table top as the table top is not designed to hold the weight of the
base and you could damage the table and the electronics that run the
tabletop.
If you need to move the table, place one person at each corner of the
table. Have each person firmly grasp the metal leg at the bottom and lift
the table from the base. As when moving any piece of heavy equipment,
bend at the knee and lift with your legs beneath you and not with your
back. Be aware and use proper, safe procedures for moving a table of
this size to prevent injury to yourself or the table. We strongly recommend
that a flatbed dolly be used that can hold the entire table if moving the
table involves any distance.
Grasp
at base
corner
•
If the seat will not move forward or back, check that the 2 fold
down sections are fully open or down all the way.
•
Check the outlet to be sure that it has power and that the power
cable is plugged in. Try the foot control and hand control to see if
anything works. If nothing works, then possibly there is a loose
connection or a faulty controller. Also, make sure the GFCI has
been
reset.
•
Unplug the power cable from the wall outlet.
•
Make sure that all male and female connectors are firmly
pushed into each other. Even a slight misconnection can
cause a possible malfunction.
• Plug the unit back into the wall and check the controls
again. If one or more of the controls still do not work,
contact Customer Service at 800-558-8850 to arrange
replacement/repair.
If the table wIll not change heIght or posItIon
cautIon
DIR
e
CTIONS
f
OR
US
e
17
DIReCTIONS fOR USe
Grasp
at base
corner