background image

3. Preparation

Inserting the Batteries

Four AAA batteries are included with your meter.

1. 

Use a Phillips screwdriver to remove the two screws holding the battery cover.
See Figure below.

2. 

Lift meter stand to expose battery cover. Remove battery cover.

3. 

Insert batteries. Follow the diagram inside the cover for correct polarity.

4. 

Replace the battery cover into its original position using the two screws removed
earlier.

6

3.1

7

Connecting the probe

The OAKTON pH/Con meter uses a specialty combination pH/conductivity/
temperature probe. The probe cable has a notched 6-pin connector to attach the
probe to the meter. 

NOTE:

Do not substitute other probes or electrodes. For a replacement probe, 
see the “Accessories” section, page 22. 

NOTE:

Keep connector dry and clean. Do not touch connector with soiled hands.

To connect the pH/Conductivity probe: 

1.

Line up the notch and 6 pins on the meter with the holes in the connector located
on the probe connector. Push down and turn the locking ring to lock into place.

See figure below.

2.

To remove probe, turn the locking ring on the probe connector. Pull probe away
from the meter.

CAUTION: Do not pull on the probe cord or the probe wires might disconnect.

3.2

locking ring

top view of 6-pin

connector

Remove these
two screws to
access battery
compartment

Battery 
compartment

Содержание 35630-00

Страница 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Страница 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Страница 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Страница 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Страница 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Страница 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Страница 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Страница 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Страница 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Страница 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Страница 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Страница 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Отзывы: