pH calibration

This instrument is capable of up to 3-point pH calibration to ensure accuracy across
the entire pH range of the meter. You can perform 1-, 2-, or 3-point calibration with
standard pH buffers 4.00; 7.00 and 10.00.

We recommend that you perform at least a 2-point calibration using standard
buffers that bracket (one above and one below) the expected sample range. You can
also perform a 1-point calibration, but make sure that the buffer value is close to the
sample value you are measuring.

This meter features three preprogrammed pH buffers (pH 4.00, 7.00 and 10.00). 
The meter automatically recognizes and calibrates to these standard buffer values,
which makes pH calibration faster and easier. See pages 22-23 for information on
our high-quality OAKTON pH buffers.

10

Calibrating for pH:

1. 

If necessary, press the MODE key to
select pH mode

. The pH indicator 

appears in the upper right hand corner 
of the display.

2. 

Rinse the probe thoroughly with 
de-ionized water or a rinse solution.

Do

not wipe the probe; this causes a build-up
of electrostatic charge on the glass surface.

3. 

Dip the probe into the calibration buffer.
The end of the probe must be completely
immersed into the sample. Stir the probe
gently to create a homogeneous sample.

4. 

Wait for the measured pH value to 
stabilize.

The READY indicator will 

display when the reading stabilizes.

See figure 

5. 

Press CAL/MEAS

to enter pH calibration

mode. The primary display will show 
the measured reading while the smaller 
secondary display will indicate the pH 
standard buffer solution. 

See figure 

NOTE: 

If using a pH buffer other than pH 7,
press the 

or 

keys to scroll up or

down until the secondary display
value is the same as your pH buffer
value (pH 4.00, 7.00 or 10.00).

A

pH

ATC

CAL

pH

pH

4.3

B

pH

ATC

READY

MEAS

C

A

B

11

pH

ATC

READY

CAL

pH

pH

ATC

pH

pH

6. 

Wait for the measured pH value to 
stabilize.

The READY indicator will 

display when the reading stabilizes.

See figure 

7. 

After the READY indicator turns on,
press ENTER to confirm calibration

A confirming indicator (CON) flashes
and disappears. The meter is now 
calibrated at the buffer indicated in the
secondary display.

See figure 

The secondary display automatically scrolls
to the next buffer calibration option. 
• If you are performing multipoint 

calibration, go to step 7. 

• If you are performing one-point 

calibration, go to step 10.

8. 

Press the

or

keys

to select the 

next buffer value you want to calibrate
(pH 4.00; 7.00 or 10.00). 

See figure 

9. 

Rinse the probe with de-ionized water
or a rinse solution, and place it in the
next pH buffer. 

10. 

Follow steps 5 to 8 for additional 
calibration points (up to 3 values). 

11. 

When calibration is complete, 
press CAL/MEAS to return to pH 
measurement mode

.

Notes

To exit from pH Calibration mode without
confirming calibration,

DO NOT press

ENTER in step 6. Press CAL/MEAS  instead.

If the selected buffer value is not within
±1.00 pH from the measured pH value:
the electrode and buffer icon blink and the
ERR annunciator appears in the lower left
corner of the display. These indicators also
flash if the buffer used is not the same as
the buffer value on the secondary display.

C

pH

ATC

READY

CAL

CON

pH

pH

ATC

READY

CAL

CON

pH

pH

pH

ATC

CAL

pH

pH

ATC

CAL

pH

pH

E

D

C

D

E

Содержание 35630-00

Страница 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Страница 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Страница 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Страница 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Страница 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Страница 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Страница 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Страница 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Страница 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Страница 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Страница 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Страница 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Отзывы: