background image

12

13

Conductivity Calibration

Ideally, you should select a conductivity standard with a value near the sample
value you are measuring. The next best method is to choose a calibration solu-
tion value that is approximately 

2

3

the full scale value of the measurement

range. For example, in the 0 to 1999 µS range, use a 1413 µS solution for calibra-
tion. Conductivity standards which are lower than 20% of the range should not
be used for calibration. It is not possible to make adjustments for values which
are in the lower 20% of the working range. See “Specifications”, page 21.

Preparing the Meter for Conductivity Calibration

You can calibrate at one point per range of the meter (a total of up to four cali-
bration points). However, only the ranges that have been calibrated have maximum
±1% full-scale conductivity accuracy. If a range was not calibrated, the meter automati-
cally detects the closest range calibrated and uses that calibration information. 

To maintain ±1% F.S. accuracy, you need to calibrate your meter at least 
once a week if: 
• you take measurements at extreme temperatures
• you are measuring in conductivity ranges lower than 100 µS

To maintain ±1% F.S. accuracy, you need to calibrate your meter at least 
once a month if:
• you are measuring in mid range solutions, and you wash the cell in deionized

water between usages. 

The conductivity of solution varies greatly with temperature. The automatic
temperature compensation (ATC) feature adjusts the fluctuations of conductivity
readings in varying temperatures to a standard temperature. These conductivity
meters feature a temperature coefficient of 2.00% per °C and normalize the 
readings at 25°C.

Calibrating for Conductivity:

1. 

Pour out two separate portions of your calibration standard and one of
deionized water into separate clean containers.

2. 

If necessary, press the MODE key to select Conductivity Mode

. The µS or

mS indicator will appear on the right side of the display.

3. 

Rinse your probe

with deionized water, then rinse the probe in one of the 

portions of calibration standard.

4. 

Immerse the probe into the second portion of calibration standard. 

The

meter’s autoranging function selects the appropriate conductivity range (four
ranges are possible). Be sure to tap the probe to remove air bubbles. Air bubbles will
cause errors in calibration.

5. 

Wait for the reading to stabilize. 

The

READY indicator lights when the reading
is stable. 

See figure

6. 

Press the CAL/MEAS key. 

The CAL indi-

cator appears above the primary display.
The primary display shows the factory
default and the secondary display shows
the temperature.

See figure 

7. 

Press the

or

keys

to scroll to the

value of your conductivity standard. Press
and hold the 

or 

keys to scroll faster. 

The meter automatically compensates for
temperature differences using a factor of
2.00% per °C.

See figure 

8. 

Press the ENTER key

to confirm 

calibration.

Upon confirmation, the CON indicator
appears briefly. The meter automatically
switches back into Measurement mode. The
display now shows the calibrated, tempera-
ture compensated conductivity value.

See figure 

9. 

For calibration in other ranges 
(maximum: four ranges) repeat steps 
1 through 9 with the appropriate 
calibration standards. See pages 
22-23 for our selection of high-quality
OAKTON conductivity calibration 
solutions.

Notes

To exit from Conductivity Calibration mode without confirming calibration, 
DO NOT press ENTER in step 8. Press CAL/MEAS  instead.

If the calibration value input into the meter is different from the factory default
value displayed by more than 30%

, the ERR annunciator appears in the lower left

corner of the display. Clean probe with alcohol. Verify that your calibration standard
is fresh and accurate.

4.4

ATC

CAL

  mS

CON

C

D

D

ATC

CAL

  mS

C

C

C

ATC

CAL

  mS

READY

C

B

ATC

READY

MEAS

  mS

C

A

B

A

Содержание 35630-00

Страница 1: ...will be made without charge OAKTON alone will determine if the product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis 13 Return of items Au...

Страница 2: ...ting 19 9 Error Messages 20 10 Specifications 21 11 Accessories 22 23 12 Warranty 24 13 Return of Items 24 2 3 pH Conductivity C Meter HOLD MODE ON OFF CAL MEAS C MEAS pH ATC pH Con 10 Series ENTER 1...

Страница 3: ...nding graphic indicator on the LCD ON OFF Powers and shuts off the meter HOLD Freezes the measured reading To activate press HOLD while in measurement mode To release press HOLD again MODE Selects the...

Страница 4: ...s a notched 6 pin connector to attach the probe to the meter NOTE Do not substitute other probes or electrodes For a replacement probe see the Accessories section page 22 NOTE Keep connector dry and c...

Страница 5: ...bration See section 4 5 on pages 14 15 for temperature calibration Preparing the Meter for Calibration The pH conductivity temperature probe included with this meter is designed for use with this mete...

Страница 6: ...ate the pH standard buffer solution See figure NOTE If using a pH buffer other than pH 7 press the or keys to scroll up or down until the secondary display value is the same as your pH buffer value pH...

Страница 7: ...containers 2 If necessary press the MODE key to select Conductivity Mode The S or mS indicator will appear on the right side of the display 3 Rinse your probe with deionized water then rinse the probe...

Страница 8: ...and the secondary display shows you what the temperature value was initially See figure 4 5 ATC CAL Temp C B A B pH ATC CAL pH pH A 4 Compare the primary display reading to a NIST traceable thermomete...

Страница 9: ...electrode has dehydrated soak it for 30 minutes in OAKTON electrode storage solution or a 2M 4M KCl solution 2 Switch on the meter The MEAS annun ciator appears on the top center of the LCD The ATC i...

Страница 10: ...a Clean probe See Probe Care Maintenance p 18 18 7 Probe Care and Maintenance Under typical operating conditions the probe will need to be replaced every 6 to 12 months In extreme applications the pr...

Страница 11: ...Error Messages LCD Indicates Cause Solution Display Err Unrecognized input Wrong input in Release key Select annunciator from keypad selected mode valid operations depending on mode CAL Err Calibratio...

Страница 12: ...D 00653 04 Electrode storage solution 1 pint bottle Keeps electrode bulb moist for faster more accurate readings Shpg wt 1 1 lbs 510 gms WD 00653 06 Electrode cleaning solution 1 pint bottle Removes b...

Отзывы: