40
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
Chapter 9 Technical Data
9.1 Specifications
FEATURES e500
Device class per MDD
II b
Classification
per
IEC60601-1
Protection against
electric shock
Class II
Protection against
electric shock
Type BF
Protection against
water
IP X4
Power Source
Compressed Oxygen, 45 to 87 PSI (3-6 Bar)
Circuit Control Source
Electric
Ventilation modes
SIMV, CPAP, Mask CPR and Intubated CPR
Tidal Volume /Rate (ml/BPM)
100/20, 150/20, 200/20, 250/20, 300/10,
350/10 400/10, 450/10, 500/10, 600/10,
650/10
± (4ml + 15%)
BTPS *
/ (± 10% or ± 1BPM)
Manually triggered Ventilation
Yes, set flow rate or pressure will be
delivered during I time then Inspiratory hold
Maximum Inspiratory hold time
6 sec.
I:E Ratio
1:2
(
±
20%)
PEEP (cm H
2
O)
0, 4, 5, 6, 7, 8, 9, 10, 12, 15, 18, 20
(± 10% or ± 2 cmH
2
O)
CPAP (cm H
2
O)
4, 5, 6, 7, 8, 9, 10, 12, 15, 18, 20
(± 10% or ± 2 cmH
2
O)
FiO2 (%)
60 or 100
(
±
15%)
Pmax (cmH
2
O)
10, 15, 20, 25, 30, 35, 40, 45, 50, 55, 60, 80
(± 10% or ± 2 cmH
2
O)
Pmin (cmH
2
O)
0, 2, 3, 4, 5, 6, 7, 8, 10, 12, 15, 20
(± 10% or ± 2 cmH
2
O)
Trigger sensitivity
3 L/min (± 10%) or 2 cmH
2
O below baseline
in CPAP mode only
APNEA back up time (sec.)
10-60
(
±
0.5s)
Battery Operating time at room
temperature (hrs.)
> 18 hrs for default settings (Data obtained
using fully charged new battery)
Altitude compensation
up to 4000m (13000 feet)
Battery Hot Swap
No
Built-in Battery charger
Yes
AC/DC power supply
100-240 VAC/ 19 VDC, 4.74 A
Patient circuit
O-Two Electronic Ventilator Circuit
Mounting Bracket
Mounting brackets for road ambulance and
mobile setting
Display 4.3"
Color
TFT
Live monitoring
Mve, Vte, Paw
(AV)
, PAW
(Peak)
, Rate (bpm),
Battery level