9.5 Electromagnetic Compatibility
O-Two e500 has been tested and complies with IEC 60601-1-2:2007
requirements.
Electromagnetic Emissions
O-Two e500 is intended for use in the electromagnetic environment specified
below. The user of O-Two e500 should assure that it is not used in
environments outside those specified:
Emission test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The O-Two e500 uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference with nearby
electronic equipment
RF emissions
CISPR 11
Class B
The O-Two e500 external power supply
is suitable for use in all establishments
including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supply buildings used for domestic
purposes.
Harmonic
emissions
IEC61000-3-2
Class A
Voltage
fluctuations/
flicker emissions
IEC61000-3-3
complies
Electromagnetic Immunity
Immunity test
IEC60601 required
test level
Actual compliance
level
Electrostatic discharge
(ESD) IEC61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
Electrical fast transient/
burst
IEC61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Surge IEC61000-4-5
± 1 kV line to line;
± 2 kV line to earth
± 1 kV line to line;
± 2 kV line to earth
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
<5%U
T
for 0.5 cycle
40%U
T
for 5 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 s
<5%U
T
for 0.5 cycle
40%U
T
for 5 cycle
70%U
T
for 25 cycle
<5%U
T
for 5 s
Power frequency (50/60
Hz) magnetic field
IEC61000-4-8
3 A/m
30 A/m