17
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
4.1.4 Connecting Patient Circuit
1. Attach the O-Two patient circuit (Figure 3) to 22mm gas output
connector (S) in Figure 2.
2. Connect the 2 sensor hoses (V) in Figure 3 of the patient circuit to their
corresponding connectors (Q) & (R) in Figure 2.
3.
DO NOT connect the patient circuit to a patient until the ventilator is
turned on (Step 4.1.5) and Start-up Selection dial (Shown in Figure 6) is
displayed.
4.1.5 Turning Ventilator ON
To turn on the ventilator, Press the ON/OFF button (N) in Figure 1 for one
second. During that second the associated green LED will start blinking at a
high frequency. After 1 second the ventilator will turn on but with no
ventilation at this point.
If the button is pressed and released for less than a second, the ventilator will
remain OFF.
Note
Do not connect the patient circuit to a patient or the test lung before turning
on the ventilator.
SIMV is the default start-up ventilation mode for e500.