46
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
Conducted RF
IEC61000-4-6
3Vrms: 50kHz to 80
MHz outside ISM
bands
3 Vrms
10Vrms: 150kHz to 80
MHz in ISM bands
10 Vrms
Radiated RF
IEC61000-4-3
10 V/m @
80 MHz to 2.5 GHz
30 V/m
O-Two e500 is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The user can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the O-
Two e500 as recommended below, according to the maximum output power
of the communications equipment:
Rated
maximum
output power
of transmitter
W
Separation distance (m)
150kHz to 800
MHz outside
ISM bands d =
3.5/3 *
√
P
150kHz to 800
MHz in ISM
bands d = 1.2
√
P
80 MHz to
800 MHz
d = 1.2
√
P
800 MHz to
2.5 GHz
d = 2.3
√
P
0.01 0.12 0.12
0.12
0.23
0.1 0.4 0.4
0.4
0.7
1 1.16 1.2
1.2
2.3
10 3.8 3.8
3.8
7.3
100 12 12
12
23