21
e500 User Manual
(15PL1011 - Rev.7 6_6_2017) English version
Press the Control Selection Knob (M) to confirm the patient size
selection and connect the patient valve to the patient, the ventilator
starts SIMV ventilation with the selected Vt and BPM. The default
parameters of this mode are listed in Table- 1.
c.
Set up the desired ventilation setting:
Health care providers may choose or change the ventilation mode or
parameter setting any time during ventilation by the following method:
Rotate the Control Selection Knob (M) and move the yellow cursor
(Figure 8) to section 3 of the screen (Figure 4) for ventilation mode
setup, or to the parameter to be set up located at section 6 of the screen
(Figure 4). The user must confirm the selection by pressing the Control
Selection Knob (M). Once confirmed, the selected area will be
highlighted with solid contrast background (Figure 9).
Navigate among the available settings by rotating the Control Selection
Knob (M). Press the Control Selection Knob (M) to choose the desired
setting. The chosen setting will turn yellow with the flashing
confirmation symbol
and the Confirmation indicator (L) to guide
users to activate the setup by pressing the Control Selection Knob (M)
again.
Press the Control Selection Knob (M) to activate the setup. The operator
can also press the Cancel button (K) to go back to previous parameters
before activation.
Note
If no selection occurs within 10 seconds or the Control Selection Knob
(M) is not pressed to confirm the changed parameter setting within 10
seconds, changes will be cancelled and the previous parameter values
will remain.
SIMV Settings
Default
Rate (BPM)
Selected or 10
Vt (ml)
Selected or 500
PEEP (cm H
2
O) 5
P max (cm H
2
O) 30
P min (cm H
2
O) 3
I:E 1:2
O
2
(%)
100%
F trig. (L/min)
3
Table-1
SIMV mode default settings