53
e700 User Manual
AC/DC Power Supply
Model
PMP90-13-2 01CV0105
Input Voltage
100 - 240 VAC
Input Current
1.06 - 0.45 A
Input Frequency
47 - 63 Hz
Output Voltage
19 VDC
Output Current
4.74 A Maximum
DC Output Plug
2.5 x 5.5 x 11 mm
DC Output Cable Length
6’
AC Power Cord Length
6’
Weight
642 g
EMC performance
Meet requirements of IEC60601-1-2
Safety Standards
Meet requirements of
IEC60601-1:2005
Note:
Upon disconnecting AC Power supply, Ventilator will
automatically switch to Battery operation without aff ecting
ventilator behaviour.
9.4. Battery Operating Time
Battery operation time ranges from 22 hours in normal operating
temperature to 16 hours under extreme low operating temperature.
9.5. Electromagnetic Compatibility
O-Two e700 has been tested and complies with IEC 60601-1-2:2007
requirements.
Electromagnetic Emissions
O-Two e700 is intended for use in the electromagnetic environment
specifi ed below. The user of O-Two e700 should assure that it is not
used in environments outside those specifi ed: