11
e700 User Manual
Note:
This feature is active in ventilation mode screens only.
3.2.7. Cancel
The Cancel key (K) (Figure 01) allows the operator to return to the
previous settings if the last unconfi rmed changes in settings are not
required.
3.2.8. Pause/Resume
During activation of the Pause/Resume key (H) (Figure 01), the
ventilator will stop ventilating with all un-locked keys kept active
except for the Manual/Inspiratory Hold key (E).
To activate the Pause function, proceed as follows:
1. Press the Pause/Resume key (H) (Figure 01). The Pause symbol
will fl ash on the left of the screen along with the confi rmation
symbol as well as the Confi rmation indicator (L) (Figure 01)
to guide users to activate the pause function by pressing the
Control Selection Knob (M) (Figure 01).
2. The symbol will fl ash for 10 seconds and then disappear if
the Control Selection Knob (M) (Figure 01) is not selected.
Users can also press the Cancel key (K) (Figure 01) to quit this
selection before the 10 seconds.
3.2.6. Screen brightness
Pressing the Screen Brightness key (F) (Figure 01) will switch
between 4 diff erent brightness levels as follows:
1. Light background with dark color text and waveform at 100%
light intensity.
2. Light background with dark colored text and waveform at 35%
light intensity.
3. Dark background with light colored text and waveform at 100%
light intensity.
4. Dark background with light colored text and waveform at 35%
light intensity.