52
e700 User Manual
When a gas source (medical oxygen) is supplied to the e700
ventilator via the gas input connection, the gas will fl ow into the
Flow control valve which is used to control both the fl ow and rate
of ventilation.
The output of this valve is connected to a selector switch which is
used to direct the fl ow path either directly to the ventilator output
(if 100% oxygen ventilation is required) or through a Venturi system
used to entrain air to provide an oxygen concentration of 60%.
9.2. Circuit Description
9.3. Battery & Power Supply
Battery Pack
Type 01 Type 02
Battery Cell Type
Rechargeable Lithium Ion Cell
Type
4ICR19/65-3
4INR19/66-3
Nominal Capacity
7500 mAh, Min 111 Wh 6000 mAh, 86.4 Wh
Nominal Voltage
14.8 V
14.4 V
Max. Charging Current
3750 mA
5000 mA
Max. Charging Voltage
16.8 ± 0.1 V
Dimension mm / in
144 x 62 x 42 / 5,66 X 2,44 x 1,65
Weight
642 g / 1,41 lbs
Test specifi cation
Meet requirements
of IEC62133:2012
Meet requirements
of IEC62133:2017
Pressurized
O
2
Ambient
Ventilator
Patient Circuit
Flow
control
Pressure/
Rate
Control
Pressure
Sensors
Flow
Sensor
Ventilation
valve
Safety
valve
Sensor
Adapter
Optional
HME
Filter
Intake
Filter
Venturi
Patient