For regular maintenance, this device only needs to be wiped gently with a soft, dry
cloth. Never immerse this device or any components in water.
•
Do not carry out repairs of any kind yourself. If a defect occurs, please contact
your local authorized distributor. Use only authorized parts and accessories.
•
Keep the surface of electrode patches clean, or else the viscidity of electrode
pad will be lower. When the surface of electrode patch is dirty, use damp
cloth to clean it and dry it before using it again.
•
Do not calibrate or maintain the instrument, and is prohibited from opening,
disassembling and maintaining the instrument arbitrarily.
•
Keep the instrument out of children’s reach.
•
Please store the therapy set in dry and ventilated places.
Troubleshooting
No display in LCD
Reason: No battery OR detach fails may cause by foreign object inside the
compartment.
Solution: Check/remove batteries, re-install batteries according to polarity.
LCD Display well but no stimulation
Reason: Two-core connecting cable not connected well OR is there any
protective film left on the pad.
Solution: Detach two-core connecting cable again OR remove protective film.
Solution: Change the location of electrode pads for Mild allergy user, or shorten
the treatment time.
*Stop to use this device and do physical check for allergy before the next use
Stimulation is weak
Reason: Pads overlapped to each other.
Solution: Separate them and re-attach to the location again.
Reason: Electrode pads get dirt / electrode pads getting loose.
Solution: Use damp cloth to clean / re-attach the pads closely to the skin
Sense of Piercing pain and skin is red
Reason: Treatment time may be too long.
Solution: Suggested treatment time is about 15 minutes.
Reason: Treatment intensity maybe too high.
Solution: Take rest for 5 mins and turn down the intensity level.
Reason: Skin allergic to electrode pads.