background image

 

Model 

TTM-6/0740 

Dimensions 

 Therapy Machine 

147mm (L) x 59mm (W) x 27.5mm (D) 

Dimensions 

 Electrode Pads 

 

83mm  x 59mm x 6mm 

Dimensions 

 Connecting Cable 

Cable Length: 1.2m 

Weight 

 Therapy Machine 

2.7oz 

Power  

2 AAA (Included) 

Number of Channels / Modes 

1 / 10 

Safety category 

Type BF  

Output Specification 

Output Intensity Level 

9 Levels 

Waveform and Shape 

Pulsed symmetric, Biphasic, Square wave 

Maximum Output Voltage (±10%) 

36.5V @ 500 Ω / 54.5V @ 2k Ω / 56.5V @ 10k Ω

 

 

 

Maximum Output Current (±10%) 

73mA @ 500 Ω / 27mA @ 2k Ω / 5.7mA @ 10k Ω

 

 

 

Net Charge (per pulse) 

Maximum Phase Charge 

Maximum Average Current 

0μC @ 500 Ω

 

Maximum Phase Charge 

11.7μC @ 500 Ω

 

Maximum Average Current 

7.4mA @ 500 Ω

 

Maximum Current Density  

0.15mA/ cm

2

 

@ 500 Ω

 

Maximum Power Density 

0.00056W/ cm

2

 

@ 500 Ω

 

Pulse Duration 

20μs

-

220μs

 

Frequency 

1 Hz-330Hz 

Default Treatment Time 

15min 

Environment 

Operations 

Temperature: About 5⁰Cto 40⁰C 

 

 

Humidity: About 15% to 93%RH 

 

Storage 

Temperature: About -

25⁰C to +70⁰C

  

 

Humidity: About 0 to 93%RH 

 

Transport 

Temperature: About -

10⁰C  to 40⁰C 

 

 

Humidity: About 15% to 93%RH 

 

Barometric Pressure for above environment: About 700hPa to 1060hPa 

To  improve  performance,  these  specifications  are  subject  to  change  without  notice.    Disposal  of  the 
device, accessories and packaging shall be carried out in accordance with national and local regulations 
for proper handling of waste. 
Calibration and Service 

  The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service 

life. It is generally recommended to have 

this device

 inspected and calibrated every 2 years to ensure 

correct functioning and accuracy. Please consult you authorized distributor. 

 

Содержание TTM-6/0740

Страница 1: ...TTM 6 0740 Low Frequency Therapy Instrument Instruction Manual Please read all instructions carefully and retain for future use ...

Страница 2: ...portant Safety Information 4 5 Battery Warning Preparation 6 Package Content Parts About Therapy Machine Set up your Device 7 Getting Start After Used Instruction of Use 8 Care and Maintenance 9 Troubleshooting EMC Declaration 10 12 Specification 13 Warranty Information 14 ...

Страница 3: ...pretation of measurements or if you have or suspect you have a medical issue Do not use this product on pregnant women children or people without ability to express their own consciousness Do not use it on patients with cardiac demand pacemaker defibrillator or implanted metallic or electronic device Precautions 1 Do not move the instrument while using or when the user wishes to change treated are...

Страница 4: ...ople with abnormally sensitive skin heart disease abnormal blood pressure malignant tumor serious cerebrovascular disease acute disease epilepsy or currently under doctor s treatment should consult a doctor before use this product 5 Do not use this product in places where there are high heat inflammables and electromagnetic radiation 6 Do not use this product while bathing or sleeping 7 Do not use...

Страница 5: ...t may cause cardiac arrhythmias 19 Stimulation should not be applied over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Technical description is contained in the instruction manual BATTERY WARNING Do not mix alkaline stand carbon zinc and rechargeable batteries nickel hydride Do not mix old and new batteries Non rechargeable batteries are no...

Страница 6: ...t mode and other parameters such as intensity through function buttons shown below on the host and know about the current status through LCD at any time Treatment Mode 5 Full automatic treatment mode and 5 manual treatment mode which satisfy to different demands Electrode Pad To be attached to bare skin to perform stimulation Connecting Cable To be used to make the connection between the machine a...

Страница 7: ...g Start Tear off the protective films on electrode pads and then paste the electrode pads onto painful areas Relax your body completely Push the power switch on the right of the host to on position and press On button to power on and start treatment It s recommended to use the instrument twice a day 15 minutes each time Caution Only use electrode pads equipped with the product Do not dispose the e...

Страница 8: ...cles of Mode 1 Mode 2 Mode 5 iv Cycles of Mode 2 Mode 1 Mode 4 and v Cycles of Mode 5 Mode 1 Mode 4 that each Mode will work for 30 seconds within each cycle Default setting as Mode 1 After each conversion of treatment mode the intensity will be back to 1st level 5 Press Manual button to choose from 5 manual treatment modes which included Mode 1 to Mode 5 After each conversion of treatment mode th...

Страница 9: ...ign object inside the compartment Solution Check remove batteries re install batteries according to polarity LCD Display well but no stimulation Reason Two core connecting cable not connected well OR is there any protective film left on the pad Solution Detach two core connecting cable again OR remove protective film Solution Change the location of electrode pads for Mild allergy user or shorten t...

Страница 10: ...chment EMC Declaration Guidance and Manufacturer s Declaration Electromagnetic Immunity this device is intended for use in the electromagnetic environment specified below The customer or the user of this device should ensure that it is used in such an environment Immunity Test Compliance Electromagnetic Environment Guidance RF emissions CISPR11 Group 1 This device used RF energy only for its inter...

Страница 11: ...uld be that of a typical commercial or hospital environment 1 kV for input output lines Surge 1 kV line to line Not applicable IEC 61000 4 5 2 kV line to earth Voltage dips short interruptions and voltage variations on power supply input lines IEC610004 11 5 UT 95 dip in UT for 0 5 cycle Not applicable 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25 cyce 5 UT 95 dip in UT for 5 sec Power ...

Страница 12: ... distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey ᵃ should be less than the compliance level in each frequency range ᵇ Interference may occur In the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnet...

Страница 13: ... 00056W cm2 500 Ω Pulse Duration 20μs 220μs Frequency 1 Hz 330Hz Default Treatment Time 15min Environment Operations Temperature About 5 Cto 40 C Humidity About 15 to 93 RH Storage Temperature About 25 C to 70 C Humidity About 0 to 93 RH Transport Temperature About 10 C to 40 C Humidity About 15 to 93 RH Barometric Pressure for above environment About 700hPa to 1060hPa To improve performance these...

Страница 14: ...No charge will be applicable for such repair or replacement SERVICE AND REPAIR If service is required for this product you should first contact Nuvomed Inc Customer Service at info nuvomed us or by calling Toll Free Number 1 866 815 4714 Monday to Friday 10am to 6pm EST NOTE Manufacturer cannot assume responsibility for loss or damage during incoming shipment As a precautionary measure carefully p...

Отзывы: