Model
TTM-6/0740
Dimensions
–
Therapy Machine
147mm (L) x 59mm (W) x 27.5mm (D)
Dimensions
–
Electrode Pads
83mm x 59mm x 6mm
Dimensions
–
Connecting Cable
Cable Length: 1.2m
Weight
–
Therapy Machine
2.7oz
Power
2 AAA (Included)
Number of Channels / Modes
1 / 10
Safety category
Type BF
Output Specification
Output Intensity Level
9 Levels
Waveform and Shape
Pulsed symmetric, Biphasic, Square wave
Maximum Output Voltage (±10%)
36.5V @ 500 Ω / 54.5V @ 2k Ω / 56.5V @ 10k Ω
Maximum Output Current (±10%)
73mA @ 500 Ω / 27mA @ 2k Ω / 5.7mA @ 10k Ω
Net Charge (per pulse)
Maximum Phase Charge
Maximum Average Current
0μC @ 500 Ω
Maximum Phase Charge
11.7μC @ 500 Ω
Maximum Average Current
7.4mA @ 500 Ω
Maximum Current Density
0.15mA/ cm
2
@ 500 Ω
Maximum Power Density
0.00056W/ cm
2
@ 500 Ω
Pulse Duration
20μs
-
220μs
Frequency
1 Hz-330Hz
Default Treatment Time
15min
Environment
Operations
Temperature: About 5⁰Cto 40⁰C
Humidity: About 15% to 93%RH
Storage
Temperature: About -
25⁰C to +70⁰C
Humidity: About 0 to 93%RH
Transport
Temperature: About -
10⁰C to 40⁰C
Humidity: About 15% to 93%RH
Barometric Pressure for above environment: About 700hPa to 1060hPa
To improve performance, these specifications are subject to change without notice. Disposal of the
device, accessories and packaging shall be carried out in accordance with national and local regulations
for proper handling of waste.
Calibration and Service
The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service
life. It is generally recommended to have
this device
inspected and calibrated every 2 years to ensure
correct functioning and accuracy. Please consult you authorized distributor.