background image

 

Immunity Test 

IEC 60601 
Test Level 

Compliance Level 

Electromagnetic 
Environment - Guidance

 

 

3 V/m  

Not applicable 

Portable and mobile RF 
communications equipment should 
be used no closer to any part of the 
models including cables, than the 
recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. 

 

150kHz to 80MHz 

 

6 Vrms in ISM 
bands 

Conducted RF 

10 V/m 

 

IEC 61000-4-6 

80MHz to 2.7GHz 

 

 

 

 

 

 

 

 

 

 

 

10 V/m 

Not applicable 

Recommended separation distance 

 

80MHz to 2.7GHz 

 

 

 

Radiated RF IEC 
61000-4-6 

385MHz-5785MHz 
Test specifications 
for ENCLOSURE 
PORT IMMUNITY 
to RF wireless 
communication 
equipment (Refer 
to table 9 of IEC 
60601-1-2:2014) 

10V/m          80MHz 
to 2.7GHz 

d=[3,5/V

1

]×P

1/2 

d=1.2×P

1/2

    

80 MHz to 800MHz  

d=2.3×P

1/2

  

  800 MHz to 2.7GHz  

 

 

 

 

 

385MHz-5785MHz 
Test specifications 
for ENCLOSURE 
PORT IMMUNITY to 
RF wireless 
communication 
equipment (Refer to 
table 9 of IEC 
60601-1-2:2014) 

Where P is the maximum output 
power rating of the transmitter In 
watts (W) according to the 
transmitter manufacturer and d Is 
the recommended separation 
distance in meters (m).    

Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey,

 

should be less than the compliance 
level in each frequency range.

Interference may occur In the 
vicinity of equipment marked with 
the following symbol:  

 

Note 1:  At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people. 
a.

 

Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless)  telephones  and  land  mobile 
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. 
If the measured field strength in the location of this device is used exceeds the applicable RF compliance level above, the 
models KTR-201, KTR-202, KTR-203 should be observed to verify normal operation. If abnormal performance is observed, 
additional measures may be necessary, such as re-orienting or relocating this device. 

b.

 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 

Содержание TTM-6/0740

Страница 1: ...TTM 6 0740 Low Frequency Therapy Instrument Instruction Manual Please read all instructions carefully and retain for future use ...

Страница 2: ...portant Safety Information 4 5 Battery Warning Preparation 6 Package Content Parts About Therapy Machine Set up your Device 7 Getting Start After Used Instruction of Use 8 Care and Maintenance 9 Troubleshooting EMC Declaration 10 12 Specification 13 Warranty Information 14 ...

Страница 3: ...pretation of measurements or if you have or suspect you have a medical issue Do not use this product on pregnant women children or people without ability to express their own consciousness Do not use it on patients with cardiac demand pacemaker defibrillator or implanted metallic or electronic device Precautions 1 Do not move the instrument while using or when the user wishes to change treated are...

Страница 4: ...ople with abnormally sensitive skin heart disease abnormal blood pressure malignant tumor serious cerebrovascular disease acute disease epilepsy or currently under doctor s treatment should consult a doctor before use this product 5 Do not use this product in places where there are high heat inflammables and electromagnetic radiation 6 Do not use this product while bathing or sleeping 7 Do not use...

Страница 5: ...t may cause cardiac arrhythmias 19 Stimulation should not be applied over swollen infected or inflamed areas or skin eruptions e g phlebitis thrombophlebitis varicose veins etc Technical description is contained in the instruction manual BATTERY WARNING Do not mix alkaline stand carbon zinc and rechargeable batteries nickel hydride Do not mix old and new batteries Non rechargeable batteries are no...

Страница 6: ...t mode and other parameters such as intensity through function buttons shown below on the host and know about the current status through LCD at any time Treatment Mode 5 Full automatic treatment mode and 5 manual treatment mode which satisfy to different demands Electrode Pad To be attached to bare skin to perform stimulation Connecting Cable To be used to make the connection between the machine a...

Страница 7: ...g Start Tear off the protective films on electrode pads and then paste the electrode pads onto painful areas Relax your body completely Push the power switch on the right of the host to on position and press On button to power on and start treatment It s recommended to use the instrument twice a day 15 minutes each time Caution Only use electrode pads equipped with the product Do not dispose the e...

Страница 8: ...cles of Mode 1 Mode 2 Mode 5 iv Cycles of Mode 2 Mode 1 Mode 4 and v Cycles of Mode 5 Mode 1 Mode 4 that each Mode will work for 30 seconds within each cycle Default setting as Mode 1 After each conversion of treatment mode the intensity will be back to 1st level 5 Press Manual button to choose from 5 manual treatment modes which included Mode 1 to Mode 5 After each conversion of treatment mode th...

Страница 9: ...ign object inside the compartment Solution Check remove batteries re install batteries according to polarity LCD Display well but no stimulation Reason Two core connecting cable not connected well OR is there any protective film left on the pad Solution Detach two core connecting cable again OR remove protective film Solution Change the location of electrode pads for Mild allergy user or shorten t...

Страница 10: ...chment EMC Declaration Guidance and Manufacturer s Declaration Electromagnetic Immunity this device is intended for use in the electromagnetic environment specified below The customer or the user of this device should ensure that it is used in such an environment Immunity Test Compliance Electromagnetic Environment Guidance RF emissions CISPR11 Group 1 This device used RF energy only for its inter...

Страница 11: ...uld be that of a typical commercial or hospital environment 1 kV for input output lines Surge 1 kV line to line Not applicable IEC 61000 4 5 2 kV line to earth Voltage dips short interruptions and voltage variations on power supply input lines IEC610004 11 5 UT 95 dip in UT for 0 5 cycle Not applicable 40 UT 60 dip in UT for 5 cycle 70 UT 30 dip in UT for 25 cyce 5 UT 95 dip in UT for 5 sec Power ...

Страница 12: ... distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey ᵃ should be less than the compliance level in each frequency range ᵇ Interference may occur In the vicinity of equipment marked with the following symbol Note 1 At 80 MHz and 800 MHz the higher frequency range applies Note 2 These guidelines may not apply in all situations Electromagnet...

Страница 13: ... 00056W cm2 500 Ω Pulse Duration 20μs 220μs Frequency 1 Hz 330Hz Default Treatment Time 15min Environment Operations Temperature About 5 Cto 40 C Humidity About 15 to 93 RH Storage Temperature About 25 C to 70 C Humidity About 0 to 93 RH Transport Temperature About 10 C to 40 C Humidity About 15 to 93 RH Barometric Pressure for above environment About 700hPa to 1060hPa To improve performance these...

Страница 14: ...No charge will be applicable for such repair or replacement SERVICE AND REPAIR If service is required for this product you should first contact Nuvomed Inc Customer Service at info nuvomed us or by calling Toll Free Number 1 866 815 4714 Monday to Friday 10am to 6pm EST NOTE Manufacturer cannot assume responsibility for loss or damage during incoming shipment As a precautionary measure carefully p...

Отзывы: