Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment - Guidance
3 V/m
Not applicable
Portable and mobile RF
communications equipment should
be used no closer to any part of the
models including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
150kHz to 80MHz
6 Vrms in ISM
bands
Conducted RF
10 V/m
IEC 61000-4-6
80MHz to 2.7GHz
10 V/m
Not applicable
Recommended separation distance
80MHz to 2.7GHz
Radiated RF IEC
61000-4-6
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY
to RF wireless
communication
equipment (Refer
to table 9 of IEC
60601-1-2:2014)
10V/m 80MHz
to 2.7GHz
d=[3,5/V
1
]×P
1/2
d=1.2×P
1/2
80 MHz to 800MHz
d=2.3×P
1/2
800 MHz to 2.7GHz
385MHz-5785MHz
Test specifications
for ENCLOSURE
PORT IMMUNITY to
RF wireless
communication
equipment (Refer to
table 9 of IEC
60601-1-2:2014)
Where P is the maximum output
power rating of the transmitter In
watts (W) according to the
transmitter manufacturer and d Is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
ᵃ
should be less than the compliance
level in each frequency range.
ᵇ
.
Interference may occur In the
vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location of this device is used exceeds the applicable RF compliance level above, the
models KTR-201, KTR-202, KTR-203 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating this device.
b.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.