Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+ 8 kV
+ 8 kV
Floors
should
be
wood,
concrete, or ceramic tile. If
floors
are
covered
with
synthetic material, the relative
humidity should be at least
30%
contact
contact
+ 2 kV,
+ 2 kV,
+ 4 kV,
+ 4 kV,
+ 8 kV,
+ 8 kV,
+ 15 kV air
+ 15 kV air
Electric fast
transient/brust
IEC 61000-4-4
+ 2 kV for power
supply lines
Not
applicable
Main power quality should be
that of a typical commercial or
hospital environment.
+ 1 kV for
input/output lines
Surge
+ 1 kV line to line
Not applicable
IEC 61000-4-5
+ 2 kV line to earth
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC610004-11
< 5% UT (>95% dip
in UT.) for 0.5 cycle
Not applicable
40% UT (60% dip in
UT.) for 5 cycle
70% UT (30% dip in
UT.) for 25, cyce
< 5% UT (>95% dip
in UT.) for 5 sec
Power
frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Magnetic fields should be at
levels characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.