background image

O

2

 Quickstick

TM

 Oxygen Analyzer 

NUVAIR                                                                                                        Page 

2

 

O

2

 Quickstick Oxygen Analyzer Operation Manual 

If you have any questions on this equipment please contact Technical Support at: 
 

Nuvair  
2949 West 5

th

 St. 

Oxnard, CA  93030 
 
Phone:  

805-815-4044 

FAX:   

805-815-4196 

Email:   

[email protected] 

 
Hours:  

Monday through Friday 
8:00 AM to 5:00 PM PST USA  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

This Operation Manual contains important safety information and should always be available to those 
personnel operating this equipment.  Read, understand, and retain all instructions before operating 
this equipment to prevent injury or equipment damage. 

 
Every effort was made to ensure the accuracy of the information contained within this manual; however, we 
retain the right to modify its contents without notice.  If you have problems or questions after reading the 
manual, stop and call 

Nuvair at 1-805-815-4044 for information.

 

Содержание O2 Quickstick

Страница 1: ...Operation Manual O2 Quickstick Oxygen Analyzer 01 11 www nuvair com...

Страница 2: ...A This Operation Manual contains important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating th...

Страница 3: ...practices Notifies people of installation operation or maintenance information which is important but not hazard related 2 System Description When using the Analyzer for diving applications with mixe...

Страница 4: ...replaceable 2 1 Preparation for Use To extend Sensor life the Analyzer is supplied with the Sensor sealed in a bag When first received please verify that the bag is intact If it is torn contact your...

Страница 5: ...he person using the gas mixture Calibration in air at sea level is suitable for many applications however the closer the oxygen content of the calibration gas is to the gas being tested the more accur...

Страница 6: ...stream to be directed at any part of your body 5 Maintenance 5 1 During the first 12 months of the warranty period the O2 Quickstick should be returned to the dealer manufacturer for repair Your O2 Qu...

Страница 7: ...the electrolyte from the Oxygen Sensor or the Sensor itself The Potassium Hydroxide chemical contained in the Sensor can cause severe injury or death If electrolyte or the Sensor is swallowed seek me...

Страница 8: ...fected area with clean fresh water for at least 15 minutes and removed contaminated clothing If stinging persists get medical attention 6 Spares Order replacement Sensor Part 9510 7 Troubleshooting Co...

Страница 9: ...on value Once you have calibrated the Analyzer for temperature and humidity using the your readings for gas analysis should be correct with no further adjustment Re calibrate Analyzer for any changes...

Страница 10: ...racy 0 1 Sensor Type Electrochemical Expected Sensor Life Room Air 36 Months Power 9V Alkaline Battery Response Time Less Than 6 Seconds to 90 of Final Value Operating Temperature 32 104 F 0 40 C Stor...

Страница 11: ...d to NUVAIR for the warranty to be registered If the warranty registration card is not received within ten 10 days of purchase the warranty will begin with the date of manufacture by NUVAIR Maintenanc...

Страница 12: ...O2 QuickstickTM Oxygen Analyzer Nuvair 2011 All Rights Reserved Phone 1 805 815 4044 Fax 1 805 815 4196 2949 West 5th Street Oxnard CA 93030 USA Email info nuvair com Web www nuvair com Revision 0111...

Отзывы: