 Quickstick Oxygen Analyzer Operation Manual 

The Nuvair O

Quickstick

 Oxygen Analyzer is designed to measure oxygen levels directly from high-pressure 

cylinders in the range 0.1-100.0% O

2

.  No sampling hoses or accessories are necessary.  The O

Quickstick

 

is designed as a quick check device to test and verify that previous readings with another analyzer are correct.  
It is a water-resistant, drop-resistant, self-contained unit designed specifically for the sport and technical diving 
industry. 

 
 
 
 
 
 
 
 

The O

Quickstick

 uses a digital display with low power consumption, which includes an on-off switch.  Power 

is provided by a standard alkaline 9-volt Battery.  The Analyzer uses an electrochemical O

2

 Sensor to measure 

O

2

 content in gases.  The Sensor is disposable and user-replaceable, with a life expectancy of up to 36 months 

depending on usage.  The Battery and Sensor are both user replaceable. 
 

2.1. Preparation for Use 

 
To extend Sensor life, the Analyzer is supplied with the Sensor sealed in 
a bag.  When first received, please verify that the bag is intact.  If it is 
torn, contact your supplier for assistance.  Prior to initial use, install the 
Sensor into the Analyzer in accordance with the Sensor Replacement 
procedure.  It is not necessary to remove the Sensor between Analyzer 
uses. 
 
The Battery must also be installed prior to initial use in accordance with 
the Battery Replacement procedure.  The Battery may be removed during 
extended periods of inactivity. 

 
3.  Calibration 

 

Oxygen Analyzers must be calibrated before each use.  Improper calibration may result in the use of 
incorrect breathing gas mixtures, which may cause serious injury or death to the person using the gas 
mixture. 

 

Calibration or use of the Analyzer with a low battery may result in inaccurate readings. Inaccurate gas 
analysis can lead to serious personal injury or death. 

 

If the Analyzer has been subjected to a recent change in ambient temperature, allow it to stabilize for 
one hour before calibration. 
 
 

 

Sensor Sealed in 

Bag 

On-Off Switch 

Содержание O2 Quickstick

Страница 1: ...Operation Manual O2 Quickstick Oxygen Analyzer 01 11 www nuvair com...

Страница 2: ...A This Operation Manual contains important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating th...

Страница 3: ...practices Notifies people of installation operation or maintenance information which is important but not hazard related 2 System Description When using the Analyzer for diving applications with mixe...

Страница 4: ...replaceable 2 1 Preparation for Use To extend Sensor life the Analyzer is supplied with the Sensor sealed in a bag When first received please verify that the bag is intact If it is torn contact your...

Страница 5: ...he person using the gas mixture Calibration in air at sea level is suitable for many applications however the closer the oxygen content of the calibration gas is to the gas being tested the more accur...

Страница 6: ...stream to be directed at any part of your body 5 Maintenance 5 1 During the first 12 months of the warranty period the O2 Quickstick should be returned to the dealer manufacturer for repair Your O2 Qu...

Страница 7: ...the electrolyte from the Oxygen Sensor or the Sensor itself The Potassium Hydroxide chemical contained in the Sensor can cause severe injury or death If electrolyte or the Sensor is swallowed seek me...

Страница 8: ...fected area with clean fresh water for at least 15 minutes and removed contaminated clothing If stinging persists get medical attention 6 Spares Order replacement Sensor Part 9510 7 Troubleshooting Co...

Страница 9: ...on value Once you have calibrated the Analyzer for temperature and humidity using the your readings for gas analysis should be correct with no further adjustment Re calibrate Analyzer for any changes...

Страница 10: ...racy 0 1 Sensor Type Electrochemical Expected Sensor Life Room Air 36 Months Power 9V Alkaline Battery Response Time Less Than 6 Seconds to 90 of Final Value Operating Temperature 32 104 F 0 40 C Stor...

Страница 11: ...d to NUVAIR for the warranty to be registered If the warranty registration card is not received within ten 10 days of purchase the warranty will begin with the date of manufacture by NUVAIR Maintenanc...

Страница 12: ...O2 QuickstickTM Oxygen Analyzer Nuvair 2011 All Rights Reserved Phone 1 805 815 4044 Fax 1 805 815 4196 2949 West 5th Street Oxnard CA 93030 USA Email info nuvair com Web www nuvair com Revision 0111...

Отзывы: