background image

O

2

 Quickstick

TM

 Oxygen Analyzer 

NUVAIR                                                                                                        Page 

5

 

O

2

 Quickstick Oxygen Analyzer Operation Manual 

 

When Analyzer calibration is performed at different atmospheric conditions than the gas being 
measured, a calibration correction value may be required.  Improper calibration may result in the use 
of incorrect breathing gas mixtures, which may cause serious injury or death to the person using the 
gas mixture. 
 

Calibration should always be performed at the same temperature and humidity conditions as the gas being 
measured.  This is not always possible, for example, in a tropical environment where dry breathing gas from a 
high-pressure Scuba cylinder will be measured after Analyzer calibration has been performed in the warm, 
humid ambient air.  Under these conditions a calibration correction value may be required, as detailed in the 
Appendix, or dry air must be used for calibration. 

 

Obtain proper training before attempting special calibration procedures.  Improper calibration may 
result in the use of incorrect breathing gas mixtures, which may cause serious injury or death to the 
person using the gas mixture. 

 
Calibration in air at sea level is suitable for many applications; however, the closer the oxygen content of the 
calibration gas is to the gas being tested, the more accurate the measurement results.  The following special 
applications require calibration procedures beyond the scope of this manual: 
 

  Analysis of gases containing greater than 50% oxygen requiring calibration with pure oxygen or 

certified calibration gas 

  Analysis of gases at altitudes above sea level requiring correction for reduced atmospheric pressure 

 

3.1. Air calibration is essential before every use and should be performed in the environment that the 

reading will be taken. 

3.2. The calibration adjustment screw is the only user adjustable control on the O

Quickstick

.  It is 

located in the recessed hole on the battery end of the Analyzer. 

3.3. The Calibrator Trim Tool provided is used to calibrate the Analyzer.  It uses a very small flat 

screwdriver blade to adjust the calibration screw. 

3.4. Expose the analyzer to fresh air for two minutes or until the reading on the display has stabilized.  

Adjust the calibration screw with the calibrator trim tool until the display reads 20.9% (see Section 4). 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Battery End 

Calibrator Trim Tool 

Sensor End 

Calibrator Trim Tool Placement 

Содержание O2 Quickstick

Страница 1: ...Operation Manual O2 Quickstick Oxygen Analyzer 01 11 www nuvair com...

Страница 2: ...A This Operation Manual contains important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating th...

Страница 3: ...practices Notifies people of installation operation or maintenance information which is important but not hazard related 2 System Description When using the Analyzer for diving applications with mixe...

Страница 4: ...replaceable 2 1 Preparation for Use To extend Sensor life the Analyzer is supplied with the Sensor sealed in a bag When first received please verify that the bag is intact If it is torn contact your...

Страница 5: ...he person using the gas mixture Calibration in air at sea level is suitable for many applications however the closer the oxygen content of the calibration gas is to the gas being tested the more accur...

Страница 6: ...stream to be directed at any part of your body 5 Maintenance 5 1 During the first 12 months of the warranty period the O2 Quickstick should be returned to the dealer manufacturer for repair Your O2 Qu...

Страница 7: ...the electrolyte from the Oxygen Sensor or the Sensor itself The Potassium Hydroxide chemical contained in the Sensor can cause severe injury or death If electrolyte or the Sensor is swallowed seek me...

Страница 8: ...fected area with clean fresh water for at least 15 minutes and removed contaminated clothing If stinging persists get medical attention 6 Spares Order replacement Sensor Part 9510 7 Troubleshooting Co...

Страница 9: ...on value Once you have calibrated the Analyzer for temperature and humidity using the your readings for gas analysis should be correct with no further adjustment Re calibrate Analyzer for any changes...

Страница 10: ...racy 0 1 Sensor Type Electrochemical Expected Sensor Life Room Air 36 Months Power 9V Alkaline Battery Response Time Less Than 6 Seconds to 90 of Final Value Operating Temperature 32 104 F 0 40 C Stor...

Страница 11: ...d to NUVAIR for the warranty to be registered If the warranty registration card is not received within ten 10 days of purchase the warranty will begin with the date of manufacture by NUVAIR Maintenanc...

Страница 12: ...O2 QuickstickTM Oxygen Analyzer Nuvair 2011 All Rights Reserved Phone 1 805 815 4044 Fax 1 805 815 4196 2949 West 5th Street Oxnard CA 93030 USA Email info nuvair com Web www nuvair com Revision 0111...

Отзывы: