 Quickstick Oxygen Analyzer Operation Manual 

4.  Operation 

4.1. Open the cylinder valve slowly until a slight hiss can be heard. 
4.2. Hold the sensor end of O

Quickstick

 (the end without the lanyard) against the valve opening so that 

the Nitrox mixture passes over the oxygen sensor. Be careful not to block the exhaust ports on the 
side. 

4.3. The reading should stabilize in 15 seconds or less and the display reading taken. 
4.4. If you cannot obtain a stable reading, check that you still have a very gentle flow of gas and repeat. If 

you still cannot obtain a stable reading see Troubleshooting section or use another analyzer. 

4.5. Note that after a few seconds of the gas flow being stopped the reading will begin to change towards 

the level in the surrounding air of 20.9% O

2

 

Never expose the Oxygen Sensor to pressures other than ambient or you may cause damage and/or 
false readings.  Damaged Sensors will not provide accurate gas analysis.  Inaccurate gas analysis can 
lead to serious personal injury or death when the gas mixture that was analyzed is used for breathing. 

 

Gas, even under moderate pressures, can cause extreme bodily harm. Never allow any gas stream to 
be directed at any part of your body. 
 
5.  Maintenance 
 

5.1. During the first 12 months of the warranty period, the O

Quickstick

 should be returned to the 

dealer/manufacturer for repair.  Your O

Quickstick

 should give you three years continuous operation 

with proper care before needing sensor replacement.  

5.2. To clean the O

Quickstick

 use a damp soft cloth. 

5.3. Although designed to be water-resistant the O

Quickstick

 should not be intentionally immersed in 

liquid or left outside unprotected. 

5.4. The O

Quickstick

 is built to resist the effects of day to day shocks and drops but remember it is a 

precision oxygen analyzer and should be looked after carefully to give long trouble free service. 

5.5. Protect the O

Quickstick

 from long periods of direct sunlight and do not subject it to high or low 

temperature extremes. 

5.6. Battery Replacement – To replace the battery, gently push and remove the end cover with thumb (the 

side with the lanyard).  See figures 1 & 2.  Gently pull the battery out from the O

Quickstick

 housing.  

Replace with new battery and reassemble. 
Be careful not to pull to hard as you can separate 
 the wires from the end cap and damage the Quickstick. 
Should you separate the wires from the end cap return 
 the Quickstick to Nuvair for repair, 
 DO NOT ATTEMPT THIS REPAIR ON YOUR OWN. 

 

 
 

 

Be sure to dispose of spent, leaking, or damaged Batteries properly, according to local

 

regulations. 

 

 

Содержание O2 Quickstick

Страница 1: ...Operation Manual O2 Quickstick Oxygen Analyzer 01 11 www nuvair com...

Страница 2: ...A This Operation Manual contains important safety information and should always be available to those personnel operating this equipment Read understand and retain all instructions before operating th...

Страница 3: ...practices Notifies people of installation operation or maintenance information which is important but not hazard related 2 System Description When using the Analyzer for diving applications with mixe...

Страница 4: ...replaceable 2 1 Preparation for Use To extend Sensor life the Analyzer is supplied with the Sensor sealed in a bag When first received please verify that the bag is intact If it is torn contact your...

Страница 5: ...he person using the gas mixture Calibration in air at sea level is suitable for many applications however the closer the oxygen content of the calibration gas is to the gas being tested the more accur...

Страница 6: ...stream to be directed at any part of your body 5 Maintenance 5 1 During the first 12 months of the warranty period the O2 Quickstick should be returned to the dealer manufacturer for repair Your O2 Qu...

Страница 7: ...the electrolyte from the Oxygen Sensor or the Sensor itself The Potassium Hydroxide chemical contained in the Sensor can cause severe injury or death If electrolyte or the Sensor is swallowed seek me...

Страница 8: ...fected area with clean fresh water for at least 15 minutes and removed contaminated clothing If stinging persists get medical attention 6 Spares Order replacement Sensor Part 9510 7 Troubleshooting Co...

Страница 9: ...on value Once you have calibrated the Analyzer for temperature and humidity using the your readings for gas analysis should be correct with no further adjustment Re calibrate Analyzer for any changes...

Страница 10: ...racy 0 1 Sensor Type Electrochemical Expected Sensor Life Room Air 36 Months Power 9V Alkaline Battery Response Time Less Than 6 Seconds to 90 of Final Value Operating Temperature 32 104 F 0 40 C Stor...

Страница 11: ...d to NUVAIR for the warranty to be registered If the warranty registration card is not received within ten 10 days of purchase the warranty will begin with the date of manufacture by NUVAIR Maintenanc...

Страница 12: ...O2 QuickstickTM Oxygen Analyzer Nuvair 2011 All Rights Reserved Phone 1 805 815 4044 Fax 1 805 815 4196 2949 West 5th Street Oxnard CA 93030 USA Email info nuvair com Web www nuvair com Revision 0111...

Отзывы: