appears on the lower line of the display only during the time that the condition exists.
When any key is pressed during the error condition, the
4-20.oUt
message appears
on the upper line of the display and the
Err Lo
message appears on the lower line of
the display. This error may be caused by the flow rate falling below the low flow rate
calibration point or the low flow rate calibration point being set too high for normal
operating conditions. Changing the low flow rate calibration point is covered in
Configuring the 4-20 mA Rate Output
section of the manual.
Calibration
Calibration of the MC-II
Plus
Flow Analyzer is a simple matter of entering the
necessary parameters for calibration into the instrument by way of the keypad. The
user friendly prompts and the ability of the MC-II
Plus
Flow Analyzer microprocessor
circuitry to calculate the divisor for volume calculation and the rate multiplier for rate
calculation make calibrating the instrument a simple process.
The steps followed to calibrate the MC-II
Plus
Flow Analyzer depend on whether
liquid or gas is being measured and the units of measure. There are four categories
of measurement:
•
Liquid Measurement Using Preprogrammed Units of Measure
•
Gas Measurement Using Preprogrammed Units of Measure
•
Liquid
Measurement
Using a Calculated Divisor and Rate Multiplier
•
Gas Measurement Using a Calculated Divisor and Rate Multiplier
The steps followed to calibrate the MC-II
Plus
Flow Analyzer for each of these
categories are outlined in the following sections.
Liquid Measurement Using Preprogrammed Units of Measure
When the liquid volume is to be expressed in barrels (BBL), gallons (GAL) or cubic
meters (M³) and the rate is to be expressed in barrels per day (BPD), gallons per
minute (GPM) or cubic meters per day (M³/D), the MC-II
Plus
Flow Analyzer
calculates the divisor and rate multiplier. The information needed to calibrate the
MC-II
Plus
Flow Analyzer is the units of measure for volume, the decimal point
setting for the volume display, decimal point setting of the meter factor, the meter
factor in pulses per gallon (
PgAL
), and units of measure for the rate. An outline of
the user prompts and the steps followed for this type of calibration are below:
1. Press the
ACCESS
key to enter the Calibrate Mode.
2. At the prompt
tot Eng
, press
INCR
to select BBL, GAL, or M³. Press
ENTER
.
3. At the prompt
tot d.P
, press
INCR
to change the decimal point position for the
volume. Press
ENTER
.
Содержание MC-II Plus Panel Mount
Страница 41: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 7 Dimensions in inches mm MC II Plus DIN Panel Mount...
Страница 44: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 10 May 2005 Flow Meter Input Wiring...
Страница 46: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 12 May 2005 External Power Wiring...
Страница 47: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 13 4 4 to 20mA Rate Output Wiring...
Страница 49: ...Appendix A1 MC II Plus WP PM Flow Analyzer May 2005 A1 15 Reset Input with Power Supply and Reset Switch...
Страница 50: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 16 May 2005 Reset Using MC II Plus Battery and Reset Switch...
Страница 52: ...MC II Plus WP PM Flow Analyzer Appendix A1 A1 18 May 2005...
Страница 60: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 8 May 2005 Remote Mount Mounting Dimensions Horizontal Pole Mount...
Страница 62: ...MC II Plus WP PM Flow Analyzer Appendix A2 A2 10 May 2005...
Страница 65: ...Appendix C MC II Plus WP PM Flow Analyzer May 2005 C 1 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 67: ...Appendix C MC II Plus WP PM Flow Analyzer May 2005 C 3 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 68: ...MC II Plus WP PM Flow Analyzer Appendix C C 4 May 2005 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 69: ...Appendix C MC II Plus WP PM Flow Analyzer May 2005 C 5 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 70: ...MC II Plus WP PM Flow Analyzer Appendix C C 6 May 2005 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 71: ...Appendix C MC II Plus WP PM Flow Analyzer May 2005 C 7 Appendix C MC II Plus WP PM Configuration Menu Flowchart...
Страница 72: ...MC II Plus WP PM Flow Analyzer Appendix C C 8 May 2005 NOTES...
Страница 78: ...MC II Plus WP PM Flow Analyzer Appendix D D 6 May 2005 NOTES...
Страница 92: ......