4.6 Lock Replacement
Disassembly:
Remove the front panel.
Remove the 2 screws located at the bottom of the lock and the top screw
Disconnect the 2 micro contacts and the magnetic sensor, as well as the power supply of the motor reducer
Slide the lock unit to the left
(For the ventilated model, it is necessary to remove the mounting bolts from the power contactor on the plate in order to
leave a space to remove the lock).
Mounting:
Connect the 2 micro contacts, the magnetic sensor, and the motor reducer.
Insert the lock unit in the location provided for this purpose.
Tighten the 3 mounting bolts with their washer.
Refit the power contactor if it was removed.
Proceed with adjusting the lock.
Replace the front panel, connecting the ground wire.