3.10 Factory Reset
After loading a new version of software microcontroller, it may be necessary to perform a factory reset, which returns the
device to factory configuration. This deletes the following data:
- Lock setting (a default is recorded)
- The imbalance calibration
- User programs
- The error counters (reset to 0)
- The serial numbers of the unit and electronic cards
Access the factory reset menu
"Warning: Return to factory settings All settings
of the user, and programs will be lost.
Continue?"
OK to continue
"Do not power off unit during reset."
Wait until the following message.
Turn off & turn on the unit.
3.11 Clear Counters
The counters are used to determine the age of the device, and the rotating accessories. In case of resetting these
counters, monitoring the life of the rotor will not be correct. A rotor may therefore be used beyond their maximum
lifespan. This causes DANGER for users and machine.
This menu is primarily used in the factory.
Access the factory reset menu
"Warning: This will erase all internal counters,
including centrifugation cycles counters, and
error counters. This should be achieved in
specific cases. Keep a record of the current
settings before proceeding. Continue?"
OK to continue