3.2 Setting the Temperature Probe
Access the temperature probe calibration menu.
The upper value “Probe Temp.” corresponds to
the physical value measured by the temperature
sensor, located in the lid.
The lower value “Displayed Temp.” is the value
displayed on the screen. This is the temperature
value of the samples in the buckets, during a
centrifugation cycle.
A correction of + /- 2°C is possible in the case of
deviation with the temperature of the sample in
the buckets.
Press the “-” or “+” keys to adjust the value.
Confirm by pressing ok.