3.4 Setting the Lock
3.4.1 Lock Operation
The lock is in the locked position when the axis of the crankshaft and the oblong hole of the bolt are
perpendicular. A marking is carried out on the visible face, corresponding to this position.
Visible face - when the front panel is removed
Locked position identification marking
Internal face - after lock unit extraction.
Perpendicular position of the 2 axes
3.4.2 Setting the Time Delay of the Lock
Access the unbalance sensitivity calibration
menu.
Locking the timer:
The motor reducer timer when closing.
Adjust with the “-” and “+” buttons up to
the position described above, in Section 3.4
Unlocking the timer:
Motor reducer time delay when closing.
Adjust with the “-” and “+” buttons until the
bolt is completely clear at the opening, in
order to let the lid strike plate escape
freely.
90°