OM0236
Page 42 of 52
February/2019
10.0 Remote Contacts
The NU-813 has several contact closures for remote sensing of various functions.
10.1 Fan Relay
The fan relay contacts are normally open and closed contact closure outputs that are activated whenever the blower key
is pressed and the blower key LED indicator is on or blinking. Contact ratings are 250 VAC maximum at 2 Amps.
10.2 Alarm Relay
The alarm relay contacts are normally open and closed contact closure outputs which are activated whenever an airflow
alarm condition occurs. Contact ratings are 250 VAC maximum at 2 Amps.
10.3 15VDC Output
The 15VDC (100mA) output is generated if the blower is actively running.