6.4.2 Panel Removal for Powder Use
If the enclosure is used to contain powdered materials, periodically the work zone panels should be removed and
cleaned. Use the following procedure for panel removal.
When removing contaminated interior panels, continue to operate the
enclosures blower to maintain negative pressure. Use appropriate
Personnel Protective Equipment (PPE) (i.e. face mask, gloves, Tyvek
gown) to mitigate exposure.
a.
Open hinged window
b.
Remove (2) knurled screws on the top of the rear panel. Lower panel and clean surfaces.
Once panel is clean, remove from enclosure.
c.
Remove top panel by pushing up and back. Allow front to swing down, then back of panel can be raised
over side brackets. Again clean surfaces as required.
d.
Clean rear plenum surfaces as well as top surfaces.
Do not clean the HEPA filter surface.
e.
Replace interior panels by reversing the process.
CAUTION
!
Содержание LabGard NU-813-300
Страница 6: ...OM0236 Page 6 of 52 February 2019...
Страница 8: ...OM0236 Page 8 of 52 February 2019...
Страница 9: ...OM0236 Page 9 of 52 February 2019...
Страница 10: ...OM0236 Page 10 of 52 February 2019...
Страница 16: ...OM0236 Page 16 of 52 February 2019...
Страница 28: ...OM0236 Page 28 of 52 February 2019...
Страница 41: ...OM0236 Page 41 of 52 February 2019 9 0...
Страница 45: ...OM0236 Page 45 of 52 February 2019...
Страница 46: ...OM0236 Page 46 of 52 February 2019...
Страница 47: ...OM0236 Page 47 of 52 February 2019...
Страница 48: ...OM0236 Page 48 of 52 February 2019...
Страница 49: ...OM0236 Page 49 of 52 February 2019...
Страница 50: ...OM0236 Page 50 of 52 February 2019...
Страница 51: ...OM0236 Page 51 of 52 February 2019...
Страница 52: ...OM0236 Page 52 of 52 February 2019...