7.6 Filter Integrity Check
The filter must be scan tested before installation into the CVE. Once installed, an internal reference leak test can be
performed to assure a proper installation was performed. Challenge aerosol (One Laskin Nozzle, PAO) should be supplied in
the rear center of the workzone and the exhaust probed above the ring. When probing, a leakage shall not exceed 0.005%
of the upstream concentration.
Internal reference concentration with one Laskin Nozzle open should be:
Laskin Nozzle Concentration Formula
# Nozzles x 135 CFM x 100 ug/L
= Challenge
Inflow (CFM)
Concentrations (ug/L)
# Nozzles x 229 CMH x 100 ug/L
= Challenge
Inflow (CMH)
Concentrations (ug/L)
Airflow FPM (m/s)
60 (.30)
80 (.41)
100 (.51)
Size
Challenge (ug/L)
300/E
93
69
56
400/E
69
52
41
600/E
46
35
28
Содержание LabGard NU-813-300
Страница 6: ...OM0236 Page 6 of 52 February 2019...
Страница 8: ...OM0236 Page 8 of 52 February 2019...
Страница 9: ...OM0236 Page 9 of 52 February 2019...
Страница 10: ...OM0236 Page 10 of 52 February 2019...
Страница 16: ...OM0236 Page 16 of 52 February 2019...
Страница 28: ...OM0236 Page 28 of 52 February 2019...
Страница 41: ...OM0236 Page 41 of 52 February 2019 9 0...
Страница 45: ...OM0236 Page 45 of 52 February 2019...
Страница 46: ...OM0236 Page 46 of 52 February 2019...
Страница 47: ...OM0236 Page 47 of 52 February 2019...
Страница 48: ...OM0236 Page 48 of 52 February 2019...
Страница 49: ...OM0236 Page 49 of 52 February 2019...
Страница 50: ...OM0236 Page 50 of 52 February 2019...
Страница 51: ...OM0236 Page 51 of 52 February 2019...
Страница 52: ...OM0236 Page 52 of 52 February 2019...