OM0236
Page 27 of 52
February/2019
7.4 Motor/Blower Replacement
7.4.1
The Motor/Blower assembly should never need any preventative maintenance. But in case of a malfunction, the
following steps should be taken. Make sure unit is decontaminated before doing any repairs on the interior of the
unit. Disconnect power or unplug unit before working with any electrical wiring. Motor/Blower removal should be
done by a qualified technician.
7.4.2
First, remove the top panel. Disconnect the electrical connections to impeller. Next remove impeller via fasteners.
(See Drawing BCD-15640, Sheet 3).
7.4.3
To install motor/blower reverse the above procedure.