OM0239
Page 22 of 72
April 2021
6.1.2
Front Panel
The control system front panel contains the following functions described in detail (see Drawing BCD-16377).
6.1.2.1 Blower Key
The blower key controls the ON/OFF power to the blower. LED above key indicates: Full green for blower on,
Blinking green for blower pending, Blinking full red for blower alarm
6.1.2.2 Hidden Key
The hidden key is located just above the blower LED indicator centered in the airflow symbol. The hidden key
is used for various functions including the blower password 3 key sequence if the option is activated.
6.1.2.3 Fluorescent Light Key
The fluorescent light key controls the on/off power to the fluorescent light.
LED above the key indicates full blue for fluorescent light on, blinking blue during auto timer duration.
6.1.2.4 Ultraviolet (UV) Light Key
The UV light key controls the on/off power to the UV light (if optionally installed).
LED above indicates full yellow for UV light on, blinking yellow during auto timer duration.
6.1.2.5 Outlet Keys
The outlet key controls the ON/OFF power to the outlets.
LED above indicates full blue for outlets on, blinking blue during auto timer duration.
6.1.2.6 Red Alarm LED
The red alarm LED will indicate any alarm condition and remain indicating until the alarm condition is cleared.
A blinking red alarm LED indicates a power interruption has occurred.
6.1.2.7 Audible Alarm Silence Key
The audible alarm silence key allows user interaction to silence a window high audible alarm for cleaning /
loading purposes. The alarm silence may last up to 15 minutes, if the alarm condition still exists, the audible
alarm will again sound.
The audible alarm silence key also is used to exit all
FlowGard™
user interaction menus.
6.1.2.8 Arrow Adjustment Keys
The arrow adjustment keys allow user interaction for various functions.
6.1.2.9 ENT Key
The ENT key is used to select item on display or accept and enter current item, function or value.
6.1.2.10 Audible Alarm
The audible alarm is provided per EN 457, tested at one frequency
to a loudness of 13 db over the cabinet’s
normal running db level. The audible alarm will activate on all alarm conditions as long as they are present.
The audible alarm will activate for 30 seconds full on, then ring back 2 seconds every 5 seconds thereafter
until the alarm condition is cleared.
Содержание LabGard ES NU-543-300S
Страница 7: ...OM0239 Page 7 of 72 April 2021 ...
Страница 9: ...OM0239 Page 9 of 72 April 2021 ...
Страница 10: ...OM0239 Page 10 of 72 April 2021 ...
Страница 11: ...OM0239 Page 11 of 72 April 2021 ...
Страница 12: ...OM0239 Page 12 of 72 April 2021 ...
Страница 16: ...OM0239 Page 16 of 72 April 2021 ...
Страница 21: ...OM0239 Page 21 of 72 April 2021 ...
Страница 35: ...OM0239 Page 35 of 72 April 2021 ...
Страница 67: ...OM0239 Page 67 of 72 April 2021 ...
Страница 68: ...OM0239 Page 68 of 72 April 2021 ...
Страница 69: ...OM0239 Page 69 of 72 April 2021 ...
Страница 70: ...OM0239 Page 70 of 72 April 2021 ...
Страница 71: ...OM0239 Page 71 of 72 April 2021 ...
Страница 72: ...OM0239 Page 72 of 72 April 2021 ...