(3) Stabilization
Autoclave sterilization is recommended.
Sterilization is required first time you use and after each patient as noted below.
This handpiece can be cleaned and disinfected with a Thermo-Disinfector.
This handpiece can be autoclavable up to Max.135°C.
1) The dirt of the handpiece surface is discarded with a brush (metal is improper).Clean handpiece using water or Isopropyl
alcohol.
2) Lubricate the handpiece. (Refer to (2) Lubrication)
3) Insert into an autoclave pouch. Seal the pouch (Only if using a vacuum autoclave)
4) Autoclavable up to max.135°
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
5) Keep the handpiece in the autoclave pouch to keep clean before use.
· Do not wash, soak, or wipe off the handpiece
with/in the oxidation potential solution (strong acid, super acid solution) or
sterilized solution.
· Do not Autoclave the equipment which with residual chemical. It may cause a peel off the plating or exert a harmful
influence to internal parts.
· Store it in a place where the handpiece is not subject to air with dust, sulfur or salinity.
· Do not touch the handpiece just after the autoclave. It is extremely hot.