(3) Stabilization

Autoclave sterilization is recommended.
Sterilization is required first time you use and after each patient as noted below.

This handpiece can be cleaned and disinfected with a Thermo-Disinfector.

This handpiece can be autoclavable up to Max.135°C.

1) The dirt of the handpiece surface is discarded with a brush (metal is improper).Clean handpiece using water or Isopropyl 

alcohol.

2) Lubricate the handpiece. (Refer to (2) Lubrication)
3) Insert into an autoclave pouch. Seal the pouch (Only if using a vacuum autoclave)
4) Autoclavable up to max.135° 

ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C. 

5) Keep the handpiece in the autoclave pouch to keep clean before use.

· Do not wash, soak, or wipe off the handpiece

 

with/in the oxidation potential solution (strong acid, super acid solution) or 

sterilized solution.

· Do not Autoclave the equipment which with residual chemical. It may cause a peel off the plating or exert a harmful 

influence to internal parts.

· Store it in a place where the handpiece is not subject to air with dust, sulfur or salinity.
· Do not touch the handpiece just after the autoclave. It is extremely hot. 

CAUTION

CAUTION

001

Содержание S-Max Series

Страница 1: ...OM C0425E Surgical Handpiece 09 03 01 M 001 001 OM C0425E S Max SG20 Ti Max Ti Max Ti Max Ti Max X SG20L X SG25L X SG93L X SG93...

Страница 2: ...G20 This handpiece is used only for the Surgical Unit with Torque Calibration System such as NSK Surgic XT Plus Using this product with other surgical units may break the handpiece or bone CAUTION Onl...

Страница 3: ...the bur drill manufacturers This may cause shank bend or brake even within the recommended speed Always keep the bur drill shank clean Debris in the chuck mechanism can prevent burs from seating prope...

Страница 4: ...long surgical burs Do not use burs longer than 26mm The shanks of longer burs may warp under Always use irrigation water during the operation Head may overheat because of the high speed rotation Even...

Страница 5: ...nting and Removing Bur Drill 1 Mounting X SG20L X SG20 X SG25L 1 Insert the bur drill into the head 2 Lightly turn the bur drill until it engages with the latch mechanism Fig 2 3 Push bur drill inward...

Страница 6: ...ool and operation procedure For X SG25L X SG93L X SG93 it can connect the Irrigation Tube to the External Spray Nozzle firmly 1 External Spray Nozzle Connect the Irrigation Tube to the External Spray...

Страница 7: ...r Only a drill with internal irrigation is used 1 Connect the ends of the Y Connector to the External Spray Nozzle and Internal Spray Nozzles refer as detailed in procedures 1 and 2 2 Connect the Irri...

Страница 8: ...of the handpiece hold firmly and spray for 2 3 seconds Check the tissue if it is clean the handpiece can be autoclaved if not spray again until clean For optimum lubrication use NSK Care 3 Plus E Type...

Страница 9: ...sert into an autoclave pouch Seal the pouch Only if using a vacuum autoclave 4 Autoclavable up to max 135 ex Autoclave for 20 min at 121 C or 15 min at 132 C 5 Keep the handpiece in the autoclave pouc...

Страница 10: ...od Optic X SG20L X SG25L X SG93 Clean the Glass Rod Optic with cotton swab soaked in alcohol Fig 12 CAUTION Do not use a pointed tool or sharp edged to clean the Glass Rod Optic They may damage the mi...

Отзывы: