(2) Removing

Push the push button with a thumb, and the bur/drill is removed. (Fig. 5)

Fig. 5

CAUTION

· Depress the head cap button until it is almost parallel to the head cap.
· If removing the bur becomes difficult, depress the Push-button, grip the bur with fine pliers and gently remove the bur. This 

case may be prevented if you often change the position of the bur in the chuck.

X-SG20L/SG20 has 3 irrigation methods are available; External, Internal or both of them, 
depending on tool and operation procedure. For X-SG25L, X-SG93L/X-SG93, it can connect 
the Irrigation Tube to the External Spray Nozzle firmly.

(1) External Spray Nozzle

Connect the Irrigation Tube to the External Spray Nozzle firmly. (Fig.6)

(2) Internal Spray Nozzle (Fig.7)

Only a drill with internal irrigation is used.

1) Set the nozzle holder.
2) Insert Irrigation Tube in Internal Spray nozzle.
3) Put the Internal Spray Nozzle into the head firmly.
4) Put the Internal Spray Nozzle into the slot on the Nozzle Holder.

Internal Spray Nozzle: 10000324

Nozzle Holder: X-SG20L: 20000396 / SG20:20000357

Fig. 7

Fig. 6

Irrigation Tube

Nozzle Holder

Irrigation Tube

Internal Spray Nozzle

Fig. 7

3. Spray Nozzle

Fig. 6

Irrigation Tube

External 

Spray 
Nozzle

001

Содержание S-Max Series

Страница 1: ...OM C0425E Surgical Handpiece 09 03 01 M 001 001 OM C0425E S Max SG20 Ti Max Ti Max Ti Max Ti Max X SG20L X SG25L X SG93L X SG93...

Страница 2: ...G20 This handpiece is used only for the Surgical Unit with Torque Calibration System such as NSK Surgic XT Plus Using this product with other surgical units may break the handpiece or bone CAUTION Onl...

Страница 3: ...the bur drill manufacturers This may cause shank bend or brake even within the recommended speed Always keep the bur drill shank clean Debris in the chuck mechanism can prevent burs from seating prope...

Страница 4: ...long surgical burs Do not use burs longer than 26mm The shanks of longer burs may warp under Always use irrigation water during the operation Head may overheat because of the high speed rotation Even...

Страница 5: ...nting and Removing Bur Drill 1 Mounting X SG20L X SG20 X SG25L 1 Insert the bur drill into the head 2 Lightly turn the bur drill until it engages with the latch mechanism Fig 2 3 Push bur drill inward...

Страница 6: ...ool and operation procedure For X SG25L X SG93L X SG93 it can connect the Irrigation Tube to the External Spray Nozzle firmly 1 External Spray Nozzle Connect the Irrigation Tube to the External Spray...

Страница 7: ...r Only a drill with internal irrigation is used 1 Connect the ends of the Y Connector to the External Spray Nozzle and Internal Spray Nozzles refer as detailed in procedures 1 and 2 2 Connect the Irri...

Страница 8: ...of the handpiece hold firmly and spray for 2 3 seconds Check the tissue if it is clean the handpiece can be autoclaved if not spray again until clean For optimum lubrication use NSK Care 3 Plus E Type...

Страница 9: ...sert into an autoclave pouch Seal the pouch Only if using a vacuum autoclave 4 Autoclavable up to max 135 ex Autoclave for 20 min at 121 C or 15 min at 132 C 5 Keep the handpiece in the autoclave pouc...

Страница 10: ...od Optic X SG20L X SG25L X SG93 Clean the Glass Rod Optic with cotton swab soaked in alcohol Fig 12 CAUTION Do not use a pointed tool or sharp edged to clean the Glass Rod Optic They may damage the mi...

Отзывы: