3-1 Cleaning (Clean-Head System)
1) Half fill a cup with clean water.
2) Brush off the debris at the Clean Head Holes. Do not use a wire brush. (Fig. 2)
3) Rotate the handpiece and immerse half of the handpiece head. (Fig. 3)
4) Rotate and stop intermittently the handpiece about 2 to 3 seconds. (About 3 times)
5) By the function of the clean head system, the clean head hole can be washed.
6) Stop the handpiece and wipe it dry. If the dirt could not remove from the hole, remove the cartridge
and clean it by brush.
7) Perform lubrication and sterilization. (3-5 Lubrication and Sterilization)
CAUTION
Use only clean water to maintain the Clean Head System.
3. Maintenances