1) Remove the handpiece from the coupling and loosen the Taper
Ring to remove. (Fig. 7)
2) Pull out the joint then remove the Non-Retraction Valve Pull out
the old Non-Retraction Valve with thin stick. (Fig. 10)
3) Insert New Non-Retraction Valve securely.
4) Mount the Joint into the handpiece. Make certain the pipe and
joint together. (Fig. 11)
5) Tighten the Taper Ring tight.
CAUTION
Make certain that the Taper Ring is firmly tightened. If the Taper Ring is loose water and air leakage could occur.
Fig. 11
Joint
Handpiece
Fig. 10
Joint
Non-Retraction Valve
Align