8
CAUTION
· Make sure that the O-Ring is removed, when removing the Cartridge. O-Ring might be remained in the Head, when removing the Cartridge. If
there is a remained O-Ring, the new Cartridge can not be inserted properly.
· Use only NSK genuine cartridge. If you use other cartridge, specification may not be satisfied and we can not warranty of this situation.
· The head cap screw thread is very fine. To prevent damage to the thread always first tighten the handpiece head cap using fingers, then
secure with the head cap wrench.
· Make sure that the O-ring (Fig. 6) is correctly located on the cartridge.
4. Replacing the Cartridge
1) Insert a dummy bur into the chuck.
2) Unscrew the Head Cap by turning counter-clockwise and remove.
3) Push up the dummy bur upwards to remove Cartridge.
4) Clean the interior of the Head casing.
5) Insert the new Cartridge.
6) Replace the Head Cap and firmly screw into place.
SX-PU03: Order Code P1140200
Fig. 6
Wrench
Cartridge
Head Cap
Head
O-Ring
PUSH