6
3-3 Cleaning (Handpiece)
This handpiece can be washed via Thermo Disinfector.
Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth. Do not use a wire brush.
Refer to the manual of Thermo-Disinfector for performing it.
3-2 Cleaning (Spray Port)
When spray nozzle are clogged, or spray does not exit evenly from three ports, clean the ports as
follows:
1) Remove the bur from the chuck.
2) Take out the cleaning wire from the back end of the brush holder. Insert the wire straight into the
spray ports with caution.
Do not forcibly insert the wire into the port
. Damaged or disoriented port
could cause the spray diverge or directed away from the bur, and the cooling efficiency deteriorates.
3) After use, clean the cleaning wire. Push into the brush holder with the pointed end of the wire inward.
Fig. 4
Cleaning Wire
Spray Ports
Содержание s-max pico BLED
Страница 1: ...High Speed Air Turbine Handpiece S Max pico OM T0496E 001...
Страница 2: ......
Страница 16: ......