ANALOG CHANNELS CONFIGURATION – CUSTOM CALIBRATION
In the "Custom Calibration" button, you can enter up to 10 custom calibration points for each analog channel. The correct
way to do this is as described below:
1. Disable the custom calibration of the channel to be adjusted, deleting all points configured for it, and apply this
setting.
2. In the points to be adjusted, apply the standard value and write down the value indicated by the equipment.
3. Afterwards, insert the pairs of Custom Calibration points (source value and indicted value) and apply the
configuration.
Analog Channels - Custom Calibration
Analog Channels - Custom Calibration
All customized calibration points should be deleted before measuring new points for another cusomtized
calibration! The existence of previous points will interfere in the measurements and the new calibration
might have errors due to this fact.
For the same reason, customized calibration points should be configured in the
FiledLogger
all at the
same time.